Dr. Schilling is the Medical Director of Imaging & Intervention, Center for Breast Care, Boca Raton Community Hospital, Boca Raton, FL. Dr. Conti is an Associate Professor of Radiology, Division of Nuclear Medicine, and the Clinical Pharmacy and Biomedical Engineering Director, PET Imaging Science Center, Los Angeles, CA. Dr. Adler is the Medical Director, Adler Institute for Advanced Imaging, Jenkintown, PA. Dr. Tafra is the Medical Director, Anne Arundle Medical Center Breast Center, Annapolis, MD. Disclosures: Dr. Schilling and Dr. Tafra are site Principal Investigators on a multicenter trial sponsored by the National Institutes of Health (NIH) and Naviscan PET Systems (San Diego, CA). Dr. Adler has done some consulting work for Naviscan. Dr. Conti is a consultant on the Scientific Advisory Board of Naviscan.
Breast cancer is one of the most frequently diagnosed cancers in women, touching the lives of roughly 213,000 women in 2006.1 In
addition to invasive breast cancer, approximately 62,000 cases of in
situ breast cancer were projected to occur in 2006, with 85% of these
being ductal carcinoma in situ. Even with these increased numbers,
mortality is decreasing. Screening mammography has been the gold
standard for breast cancer surveillance for 3 decades and is credited
with decreasing mortality by 33%.2 Because mammography has
resulted in earlier detection of breast cancer, more patients today are
candidates for, and choose, breast-conserving surgery instead of
mastectomy. However, mammography is frequently inadequate as a planning
tool for lumpectomy. In addition, it has very limited value in women
with dense breasts. This may explain the finding that residual cancer can
be found in as many as 30% to 60% of patients after lumpectomy,
resulting in a second trip to the operating room.3
One
goal of molecular imaging is to preoperatively identify which patients
are best served by lumpectomy and define the margins for surgery by
identifying metabolic abnormalities in tissue, potentially decreasing
the number of second surgeries needed for resection of residual disease.
Any technology that could more precisely map the extent of both
invasive and noninvasive disease would lead to more precise surgery.
This article will review the history of molecular imaging in breast
cancer with a special emphasis on the use of the new technology positron
emission mamography (PEM), in pre- and postoperative breast cancer
management.
Scintimammography
Nuclear imaging of
breast cancer is not a new approach. Thallium 201 (Tl-201) was first
shown to be an effective breast tumor-avid agent in 1978.4 Waxman et al5 later
compared Tl-201 with technetium (Tc)-99m sestamibi in 1993 and reported
that sestamibi had a higher sensitivity in detecting breast cancer. In
1994, Kao et al6 used Tc-99m sestamibi to examine palpable
breast masses. Of the 32 cases of breast cancer found, 27 were detected
by Tc-99m sestamibi, with a sensitivity of 84%, a specificity of 100%,
and an accuracy of 87%.6 This was a small but promising
result that prompted 2 large multicenter studies to evaluate
scintimammography in more than 2500 patients. These studies found a
sensitivity and specificity of scintimammography in the detection of
malignant breast tumors of approximately 85%.7,8 Further
analysis revealed that sensitivity fell significantly with decreasing
tumor size. This was further investigated by Tofani et al,9 who
found that scintimammography had a sensitivity of only 48% for tumors
≤1 cm in size. It is important to note that all of these initial studies
were completed using conventional whole-body (WB) gamma cameras; during
this scanning, the patient is recumbent on an imaging table.
More
recently, it was thought that the sensitivity and specificity of
sestamibi imaging could be increased with the use of a small
field-of-view (SFOV) gamma camera in which the detector is in close
proximity to breast tissue. Recent results using this type of camera
were reported by the Mayo clinic group.10 In this 100-patient
study, they found a detection sensitivity of 29% for tumors <5 mm,
86% for tumors between 6 and 10 mm, and 97% for tumors >11 mm in
diameter. The investigators thought that the technique may be an
important adjunct to mammography and that greater sensitivity might be
achieved with the optimization of collimation and the addition of a
second detector. This evolution of technology from WB imaging to SFOV or
organ-specific imaging in scintimammography was also seen with positron
emission tomography (PET), as described in the next section. The
characteristics of WB versus SFOV cameras for scintimammography and PET
are compared in Table 1.
Whole-body PET in breast cancer imaging
Whole-body
PET and PEM use the radiopharmaceutical 2-deoxy-2-[18F] fluoro-D-glucose
(FDG), a positron-emitting analog of glucose, to detect metabolic
alterations within cells. This radiotracer takes advantage of the fact
that malignant cells express higher levels of the glucose transport
protein GLUT-1 and have increased glucose needs. Since 18F-FDG is taken
up like glucose but is not metabolized, it becomes metabolically trapped
within the cancer cell. The first study to use WB PET imaging for breast
cancer was published in 1989.11 This study evaluated 17
patients with metastatic disease who had primary tumors >5 cm in
size. They reported that 14 (82%) of the tumors were detected with the
18F-FDG. The positive results of this study prompted Wahl et al12 in
1991 to evaluate the efficacy of WB PET imaging for less advanced breast
cancer. They reported that FDG PET identified 10 of 10 primary breast
cancers >3 cm. Although the use of PET has been shown to be sensitive
and effective for the detection of advanced breast cancer and distant
metastatic disease in numerous studies,13,14 it is known to have limitations in detecting small, well-differentiated tumors and in-situ lesions.15 Summarizing
a number of studies, the detection rate of WB PET in primary breast
cancer has a sensitivity range of 80% to 100% and a specificity range of
75% to 100%, with an accuracy of 70% to 97%, a positive predictive value
(PPV) of 81% to 100%, and a negative predictive value (NPV) of 52% to
89%.12,16-21 As would be expected from prior study results,
the reports of the highest sensitivities for breast cancer detection
have been studies that included patients with large tumors.
A recent study by Mavi et al22 examined
the use of dual–time-point imaging as a way to increase sensitivity of
WB PET in primary breast cancer detection. In general, this approach
involves measuring the FDG uptake in a lesion at 2 separate times, such
as 60 and 120 minutes after tracer injection. Cancer cells typically
show continued FDG accumulation over this time interval, while normal or
nonmalignant cells typically do not. Mavi et al22 reported
detection sensitivities of 90.1% for lesions >10 mm (excluding 2
indeterminate cases), 82.7% for lesions 4 to 10 mm (excluding 1
indeterminate case), and 76.9% for ductal carcinoma in situ (DCIS)
(excluding 4 indeterminate cases). Thus, although the authors concluded
that dual–time-point imaging can improve the sensitivity and accuracy of
18F-FDG in assessing patients with primary breast cancer, there still
appears to be room for improvement. Finally, this approach does not
address the need for greater accuracy in presurgical planning.
PEM in breast cancer imaging
High-resolution
PEM is a new technology that is designed for the imaging of specific
small body parts where high-resolution detection of FDG uptake is
needed. An available dual-detector system (Naviscan PET Systems, Inc.,
San Diego, CA) consists of 2 flat, high-resolution detector heads mounted
directly to compression paddles that can be rotated to optimize
imaging, such as in acquiring mediolateral oblique and craniocaudal
breast views. By lightly compressing the breast tissue during
acquisition, the image can be acquired in positions that are analogous
to those used in mammography, which allows for image coregistration and
comparison. The close proximity of the 13mm crystal detectors and
limited angle tomographic reconstruction results in an in-plane spatial
resolution of 1.5 mm full width at half maximum, compared with the 4.2
to 6.5-mm axial resolution found in commercially available WB PET
scanners. The difference in image resolution between WB PET imaging and
PEM can be seen in Figure 1. These images were acquired using a phantom
filled with ~100 uCi 18F-FDG imaged on the high-resolution PEM system
(standard 10-minute acquisition) and a WB PET scanner using a
“Brain-mode” 356 × 356 field of view matrix, 10-minute acquisition
(Biograph PET scanner, Siemens Medical Solutions USA, Inc., Malven, PA).
It is important to remember that the spatial resolution in normal
breast tissue in the WB PET scanner would be negatively impacted by
recumbent imaging because of normal respiratory motion that was not
replicated in this phantom WB PET acquisition. Respiratory motion is not
an issue with PEM imaging, as the breast is immobilized (Table 1).
The results of the first PEM pilot study in breast cancer, which used a 10-mm crystal, were published in 2005.23 Of
the 44 women with confirmed breast cancers, 39 of the 44 primary index
tumors were seen. In addition, of the 19 patients who were undergoing
breast-conserving surgery, PEM correctly predicted 75% of patients with
positive margins and 100% with negative margins. Figure 2 provides an
example of a PEM image that highlights its utility in detecting
multicentric disease. Positron emission tomography also detected 4 of 5
histologically proven incidental breast cancers, 3 of which were not
seen by any other imaging modality. One such example case is shown in
Figure 3. The authors concluded that PEM showed promise in detecting
breast malignancies and assisted in planning breast-conserving surgery.
The
results of a second, larger multi-center study that examined the
performance efficacy of the PEM in women with known breast cancer or
suspicious mammography findings were published in 2006.24 In
nondiabetic patients with proven breast cancer, PEM was found to have a
cancer detection sensitivity of 91%, specificity of 93%, NPV of 88%, and
an accuracy of 92%. Most importantly, PEM accurately identified 91% of
the cases of DCIS preoperatively. In this study, 36 of 73 biopsies (49%)
prompted by conventional imaging alone proved to be benign; however,
combining conventional imaging with PEM resulted in few false positives,
with a PPV of 95%. This finding highlights the advantage of combining
anatomic and metabolic characterization in cancer detection.
The
advantage of highly sensitive metabolic imaging is further highlighted
in a patient in whom DCIS was seen only with the PEM imaging but missed
with magnetic resonance imaging (MRI) and WB PET (Figure 4). This
patient was part of a PEM, MRI, WB PET trial being undertaken by
Schilling.25 Preliminary findings of this ongoing trial were presented at the Society of Nuclear Medicine meeting in June 200725 and
are summarized in Table 2. In addition to the benign findings listed,
pathologic analysis identified 39 distinct cancerous or in situ cancerous
lesions in the 28 study subjects. With a sensitivity of 92.3%, positron
emission mammography had the greatest sensitivity, while WB PET had a
sensitivity of only 39%. At the current interim analysis of this small
study, the results suggest that PEM is superior to WB PET and is at
least as sensitive as MRI in identifying malignant breast disease. In
addition, the author has found that the metabolic imaging provided by
PEM is helpful in premenopausal women being evaluated for breast cancer
because the lesion-to-back-ground FDG uptake ratio does not appear to be
affected by hormonal changes in the menstrual cycle. Like WB PET,
however, FDG is not recommended for use in pregnant women.
In an attempt to improve the detection sensitivity and specificity of PEM further, Adler et al26 investigated
the application of dual–time-point imaging using PEM in a pilot study
of 11 patients. They found a median increase in lesion-to-background
ratio of 36% (range 16% to 85%). This improved ratio appeared to be due
to a reduction in mean background FDG levels (Figure 5). Of interest was
the finding that 3 of 3 benign lesions showed a decrease in
lesion-to-background ratio at the second measurement. These results are
promising and suggest that delayed image comparison may prove helpful in
discriminating benign from malignant lesions. Additional, large-scale
studies will be needed to test this hypothesis.
PEM and WB PET/CT in surgical planning and breast cancer recurrence
An
ideal goal for any molecular imaging approach would be to provide a map
of the extent of both invasive and noninvasive disease to assist the
surgeon in undertaking more precise excision of involved breast tissue
and to more accurately monitor for recurrence. An advantage of the PEM
technology is that it uses mammographic positioning, which allows for
direct correlation of PEM images with mammography for both initial and
recurrence imaging (Figure 6). Positron emission mammography can also
provide a tomographic image that may further assist the surgeon in
determining the ideal approach to ensure negative margins. Another
molecular imaging goal would be to provide assistance in determining the
extent of disease (eg, lymph node involvement). While it is too early
to determine the sensitivity and specificity of the PEM technology for
this goal, some commercial sites are obtaining useful images of lymph
node involvement (Figure 7).
Approximately 16% to 30% of patients
with local-regional recurrence have been found to have metastatic
disease when evaluated with WB 27-30 In addition, 24% of patients with breast cancer recurrence locally will develop a site of distant disease within 18 months.29 Although
this highlights the importance of WB PET in the detection of metastatic
disease, an important question remains: whether the use of PEM imaging
for routine surveillance in women treated for breast cancer might allow
detection of local-regional recurrence at an earlier stage. The
corollary is, if detected earlier, could earlier therapy prevent or
decrease the incidence of cancer metastasis?
Whole-body PET has
been used to predict response to neoadjuvant chemotherapy in women with
advanced breast cancer. It has been shown to have a sensitivity range of
80% to 90% and a specificity range of 50% to 80% in predicting
pathological response (eg, the decrease in FDG uptake).31,32 The wide range of values reflects the differences in the definition of response among
different centers as well as the diminished sensitivity of WB PET in
smaller lesions. Although it appears that PET can frequently determine
an early response before other forms of imaging,33 the
question of whether PEM imaging will offer an even greater advantage in
the management of neoadjuvant chemotherapy patients will need to await
the results of a clinical trial.
New molecular imaging agents for PEM
Although
the introduction of 18F-FDG in PET imaging has changed patient
management in a variety of cancers (including breast), the ultimate goal
in molecular imaging is to image the in vivo cancer biology of an
individual to allow therapy to be personalized. The introduction of new
positron-emitting imaging agents such as the cell proliferation markers
[F-18]-fluoro-L-thymidine (FLT)34 and F-18 or C-11-2'-fluoro-5-methyl-1-beta-d-arabinofuranosyluracil (FMAU),35 and [F-18]-fluoromisonidazole,36 a
radiotracer marker for tumor hypoxia, offers new opportunities for
evaluating breast cancer and might help achieve this goal. In addition,
16a-[F-18]-fluoroestradiol-17b (FES)37 ap-pears to be a
promising estradiol analog and provides an imaging approach to monitor
and predict clinical response to hormonal therapy in vivo. Positron
emission mammographic technology should be equally effective at imaging
the localization of these radiotracers, as it has been for 18F-FDG. The
current data would suggest that positron radiotracer development and
PET/PEM imaging technologies are in their infancy; however, combined,
they are bringing us closer to personalized cancer therapy.
Acknowledgment
The
authors would like to thank Ms. Mindy McKinstry of EPIC Imaging,
Portland, OR, for the axillary lymph node imaging and Judith E.
Kalinyak, MD, PhD, Medical Director, Naviscan PET Systems, Inc., for
providing assistance in preparing this manuscript.
REFERENCES
- Cancer Facts and Figures 2006. Atlanta, GA: American Cancer Society; 2006.
- Mammography Accounting Office. GAO/HEHS-96-17. Initial impact of new
federal law has been positive. Report to Congressional Committees,
Washington, DC; U.S. Government Accounting Office; October 27, 1995.
- Gwin
JL, Eisenberg BL, Hoffman JP, et al. Incidence of gross and microscopic
carcinoma in specimens from patients with breast cancer after
re-excision lumpectomy. Ann Surg. 1993;218:729-734.
- Hisadi K, Tonami N, Miyamae T, et al. Clinical evaluation of tumor imaging with 201 Tl chloride. Radiology. 1978;129:497-500.
- Waxman
AD, Ramanna L, Memsic LD, et al. Thallium scintigraphy in the
evaluation of mass abnormalities of the breast. J Nucl
Med. 1993;34:18-23.
- Kao CH, Wang SJ, and Liu TJ. The use of
technetium-99m methoxyisobutylisonitrile breast scintigraphy to evaluate
palpable breast masses. Eur J Nucl Med. 1994;21:432-436.
- Taillefer
R. The role of 99mTc sestamibi and other conventional
radiopharmaceuticals in breast cancer diagnosis. Semin Nucl
Med. 1999;29:16-40.
- Waxman AD. The role of (99m)Tc methoxyisobutylisonitrile in imaging breast cancer. Semin Nucl Med. 1997;27(1):40-54.
- Tofani
A, Sciuto R, Semprebene A, et al. 99Tcm-MIBI scintimammography in 300
consecutive patients: Factors that may affect accuracy. Nucl Med
Commun. 1999;20:1113-1121.
- O’Connor MK, Phillips SW, Hruska CB,
et al. Molecular breast imaging: Advantages and limitations of a
scintimammographic technique in patients with small breast tumors.
Breast J. 2007;13(1):3-11. Comment in: Breast J. 2007;13(1):1-2.
- Minn H, Soini I. [18F]fluorodeoxyglucose scintigraphy in diagnosis
and follow up of treatment in advanced breast cancer. Eur J Nucl
Med. 1989:15: 61-66.
- Wahl RL, Cody RL, Hutchins GD, Mudgett EE.
Primary and metastatic breast carcinoma: Initial clinical evaluation
with PET with the radiolabeled glucose analogue
2-[F-18]-fluoro-2-deoxy-D-glucose. Radiology. 1991;179:765-770.
- Schirrmeister
H, Kühn T, Guhlmann A, et al. Fluorine-18 2-deoxy-2-fluoro-D-glucose PET
in the preoperative staging of breast cancer: Comparison with standard
staging procedures. Eur J Nucl Med. 2001;28:351-358.
- Rosé C, Dose J, Avril N. Positron emission tomography for the diagnosis of breast cancer. Nucl Med Commun. 2002;23:613-618.
- Eubank WB, Mankoff DA. Evolving role of positron emission tomography in breast cancer imaging. Semin Nucl Med. 2005;35:84-99.
- Scheidhauer
K, Walter C, Seemann MD. FDG PET and other imaging modalities in the
primary diagnosis of suspicious breast lesions. Eur J Nucl Med Mol
Imaging. 2004;31(suppl 1):S70-S79.
- Adler LP, Crowe JP, al-Kaisi
NK, Sunshine JL. Evaluation of breast masses and axillary lymph nodes
with [F-18] 2-deoxy-2-fluoro-D-glucose PET. Radiology. 1993;187:743-750.
- Hoh
CK, Hawkins RA, Glaspy JA, et al. Cancer detection with whole-body PET
using 2-[18F]fluoro-2-deoxy-D-glucose. J Comput Assist Tomogr. 1993;17:582-589.
- Avril N, Dose J, Jänicke F, et al. Metabolic
characterization of breast tumors with positron emission tomography
using F-18 fluorodeoxyglucose. J Clin Oncol. 1996;14:1848-1857.
- Avril
N, Rosé CA, Schelling M, et al. Breast imaging with positron emission
tomography and fluorine-18 fluorodeoxyglucose: Use and limitations. J Clin
Oncol. 2000;18:3495-3502.
- Scheidhauer K, Schari A, Pietrzyk U,
et al. Quantitative [18F]FDG positron emission tomography in primary
breast cancer: Clinical relevance and practicability. Eur J Nucl Med.
1996 23:618-623.
- Mavi A, Urhan M, Yu JQ, et al. Dual time point
18F-FDG PET imaging detects breast cancer with high sensitivity and
correlates well with histologic subtypes. J Nucl Med. 2006;47:1440-1446.
- Tafra L, Cheng Z, Uddo J, et al. Pilot clinical trial of
18F-fluorodeoxyglucose positron-emission mammography in the surgical
management of breast cancer. Am J Surg. 2005;190:628-632.
- Berg
WA, Weinberg IN, Narayanan D, et al. High-resolution fluorodeoxyglucose
positron emission tomography with compression (“positron emission
mammography”) is highly accurate in depicting primary breast
cancer. Breast J. 2006;12(4):309-323.
- Schilling KJ. High-resolution positron emission mammography in breast cancer. J Nuc Med. 2007; 48 (suppl 2):139P.
- Adler,
LP, Narayanan D, Gammage L, et al. Quantitative improvement in breast
lesion detectability on delayed images using high resolution positron
emission mammography. J Nuc Med. 2007;48 (suppl 2):369P.
- Danforth
DN Jr, Aloj L, Carrasquillo JA, et al. The role of 18F-FDG-PET in the
local/regional evaluation of women with breast cancer. Breast Cancer Res
Treat. 2002;75(2):135-146.
- Goerres GW, Michel SC, Fehr MK, et
al. Follow-up of women with breast cancer: Comparison between MRI and
FDG PET. Eur Radiol. 2003;13: 1635-1644.
- van Oost FJ, van der
Hoeven JJ, Hoeckstra OS, et al. Staging in patients with locoregionally
recurrent breast cancer: Current practice and prospects for positron
emission tomography. Eur J Cancer. 2004; 40:1545-1553.
- Weir L,
Worsley D, Bernstein V. The values of FDG positron emission tomography
in the management of patients with breast cancer. Breast J. 2005;
11:204-209.
- Kim SJ, Kim SK, Lee ES, et al. Predictive value of
[18F]FDG PET for pathological response of breast cancer to neo-adjuvant
chemotherapy. Ann Oncol. 2004;15:1352-1357.
- Smith IC, Welch AE,
Hutcheon AW, et al. Positron emission tomography using
[(18)F]-fluorodeoxy-D-glucose to predict the pathologic response of
breast cancer to primary chemotherapy. J Clin Oncol. 2000;18:1676-1688.
- Chen
X, Moore MO, Lehman CD, et al. Combined use of MRI and PET to monitor
response and assess residual disease for locally advanced breast cancer
treated with neoadjuvant chemotherapy. Acad Radiol. 2004;11:1115-1124.
- Shields
AF. Positron emission tomography measurement of tumor metabolism and
growth: Its expanding role in oncology. Mol Imaging Biol. 2006;
8(3):141-150.
- Conti PS, Alauddin MM, Fissekis JR, et al.
Synthesis of 2'-fluoro-5-[11C]-methyl-1-beta-D-arabinofuranosyluracil
([11C]-FMAU): A potential nucleoside analogue for in vivo study of
cellular proliferation with PET. Nucl Med Biol. 1995;22:783-789.
- Alauddin
MM, Conti PS, Fissekis JD. Synthesis of [18F]-labeled
2'-deoxy-2'-fluoro-5-methyl-1-beta-D-arabinofuranosyluracil. J Label
Compd Radiopharm. 2002;45:583-590.
- Kumar P, Mercer J, Doerkson
C, et al. Clinical production, stability studies and PET imaging with
16-alpha-[18F]fluoroestradiol ([18F]FES) in ER positive breast cancer
patients. J Pharm Pharm Sci. 2007;10:256S-265S.