Dr. Schoenberg
is an Associate Professor of Radiology, an Associate Chair for
Clinical Operations, and the Section Chief, Magnetic Resonance
Imaging;
Dr. Zech
is a Fellow of Radiology and Vice Section Chief, Abdominal
Imaging;
Dr. Panteleon
is a Resident; and
Dr. Michaely
is a Resident; Institute of Clinical Radiology, University
Hospitals-Grosshadern, Ludwig-Maximilians-University, Munich,
Germany.
Dr. Reiser
is a Professor of Radiology and the Chairman, Institute of
Clinical Radiology, University Hospitals-Grosshadern and
Innenstadt, Ludwig-Maximilians-University, Munich, Germany.
Dr. Finn
is Professor of Radiology, Cardiovascular Research, Department of
Radiology, David Geffen School of Medicine at the University of
California,Los Angeles, CA.
Currently, the challenges for abdominal magnetic resonance
imaging (MRI) are confined within a magic triangle of spatial
resolution, signal-to-noise ratio (SNR), and artifacts from
respiratory motion or vascular pulsation. Within the last few
years, a number of strategies have been developed to overcome these
limitations. The use of parallel acquisition techniques on MRI
systems with multiple receiver channels now allows the increase of
spatial resolution at almost no cost in scan time.
1,2
Motion artifacts can successfully be avoided by the application of
navigator-based correction techniques, thereby synchronizing the
image acquisition with the position of the diaphragm.
2,3
Lastly, SNR constraints can now be effectively overcome by the
introduction of 3T MRI systems into the clinical routine. To truly
assess the additional value of 3T systems for abdominal MRI, a fair
comparison must be made to current state-of-the-art 1.5T MRI
systems. This way it is possible to evaluate the benefit of 3T in
view of the higher hardware costs and the substantial increase in
technical difficulties.
At the present time, state-of-the-art 1.5T MRI scanners are
equipped with at least 8 to 32 independent receiver channels,
allowing the use of parallel acquisition techniques (PATs) in all 3
directions with acceleration factors of up to 6. For PATs,
essentially 2 different algorithms are applied, which perform the
reconstruction either in the image domain (sensitivity-encoded
[SENSE]-based algorithms)
4
or k-space domain (simultaneous acquisition of spatial harmonics
[SMASH]-based algorithms).
5-8
Dependent on the vendor-specific implementation, the information of
the coil sensitivity profiles is either acquired within a SENSE
sequence or by acquisition of additional reference lines within the
same scan (modified SENSE [mSENSE] and generalized autocalibrating
partially parallel acquisition [GRAPPA]).
9
The latter implementation is known as an integrated parallel
acquisition technique (iPAT).
Current state-of-the-art imaging at 1.5T
With state-of-the-art 1.5T scanners, a slice thickness of 4 to 6
mm for abdominal imaging can routinely be applied with
2-dimensional sequences and 2 to 4 mm with 3-dimensional (3D)
sequences.Acqui-sition of a (320)
2
matrix at a (350 mm)
2
field of view allows obtaining an in-plane resolution of
approximately 1 mm
2
routinely. By combining PATs with prospective navigator correction
(eg, in T2- weighted [T2W] turbo spin-echo [TSE]) sequences, the
image acquisition of larger organs, such as the liver, is feasible
in 3 breath-holds with a breath-hold time of only 13 seconds.
2
For free-breathing acquisitions, the total acquisition time can be
effectively reduced by PATs to approximately 4 minutes. Studies
have shown that the use of PAT does not result in a substantially
larger number of artifacts from aliasing, while at the same time,
motion artifacts are decreased because of the shorter overall scan
times.
2
With these techniques, liver lesions can now be reliably detected
with T2W sequences at a size of 4 to 5 mm or less after the
administration of superparamagnetic iron oxides (SPIO). It has been
shown that the combination of dynamic gadolinium-enhanced imaging
using thin-slice 3D T1-weighted (T1W) gradient-echo (GRE) sequences
with SPIO imaging further improves the detection of small lesions
(<10 mm).
10
Higher PAT acceleration factors can be applied in sequence types
that have an intrinsically high SNR, such as steady-state
free-precession [SSFE] techniques (eg, true fast imaging with
steady-state precession [TrueFISP], balanced fast-field echo [FFE],
and fast imaging employing steady-state acquisition [FIESTA]).
These techniques are particularly favorable for single-breath-hold
3D imaging of the small and large bowel after intraluminal water
distension with acceleration in 2 phase-encoding directions. A
total PAT factor of 6 allows the acquisition of an entire data set
with 1.5-mm isotropic voxel length in a single breath-hold of
<20 seconds. This data can be individually reformatted to show
the exact extent of small or large bowel pathology.
Currently, pancreatic imaging can also be performed at very high
resolutions by the use of PAT, (320)
2
to (384)
2
matrix size, and a slice thickness of 4 mm after the administration
of manganese-containing contrast agents such as Teslascan (Amersham
Health, Ismaning, Germany).
11
Similarly, the advent of PAT has revolutionized the use of
contrast-enhanced 3D MR angiography (MRA) in the abdomen. Voxel
sizes of <1 mm
3
are now achievable within a single breath-hold.
12-14
These isotropic data sets can also be reformatted in any plane,
which offers the possibility to assess the area stenosis rather
than only the diameter stenosis. These area stenosis measurements
have shown good correlation to intravascular ultrasound in
preliminary studies on renal arteries and appear to be
significantly more accurate in terms of the measurement of the true
degree of stenosis.
14
Capabilities at 3T
As already outlined, the current achievements at 1.5T apply a
high standard in abdominal MRI. Thus, the initial step at 3T was to
reproduce similar image parameters as compared with 1.5T. The
problem is that there are a number of challenges for abdominal MRI
at 3T despite the obvious advantage of a higher SNR. First, the
specific absorption rate (SAR) increases quadratically with field
strength, resulting in a major limitation, particularly for turbo
spin-echo sequences.
15
The shorter wavelength at 3T results in substantial dielectric
resonance effects, causing significant signal inhomogeneities due
to eddy currents in the tissue, with the result of focal-field
cancellation.
15
Relaxation times are substantially longer, which necessitates the
adjustment of several imaging parameters (eg, repetition time [TR],
echo time [TE]).
Therefore, the second step for abdominal imaging at 3T has been
to effectively address these limitations by new technical
strategies. The signal inhomogeneities due to the dielectric
resonance effects can be substantially reduced by the application
of external gel pads placed on the surface of the patient. These
radiofrequency (RF) homogenization pads are not visible in the
images, nevertheless they alter the dielectric properties of the
imaging volume, improving signal homogeneity.
16
The SAR can be effectively reduced by new techniques (such as the
hyperecho technique and the variable flip angle technique)
17-19
or parallel imaging.
20
In the third step, these new techniques have been effectively
implemented on clinical scanners to further push the limits. At the
present, matrix sizes of (512)
2
and slice thicknesses of <4 mm appear feasible, while at the
same time breath-hold scans of <12 seconds or total scan times
for free-breathing acquisitions of <2 minutes can be achieved
using PAT.
Image optimization
The following sections discuss the individual steps for the
optimization of image acquisition for abdominal 3T MRI.
Optimization of image contrast
As the T1 relaxation time increases with the field strength
B
0
(with a typical increase of 30% to 50% of T1 at 3T compared with
1.5T), the sequence parameters-particularly the TR-has to be
adapted.
21
The effects on the T2 relaxation are less prominent; T2 is
effectively unchanged at 3T, but T2* is shortened.
22
The adjustment of the TR strongly depends on the flip angle.
Therefore, GRE sequences with small flip angles might require only
minor changes in TR compared with sequences with a full 90° flip
angle.
21
In daily practice at our institution, we choose a decreased flip
angle of only 50° for T1W GRE sequences because this flip angle has
been shown to produce optimal image contrast with no adaptations of
the TR. Figure 1 depicts an example of the excellent image quality
in this T1W sequence.
Changes of TEs for in-phase/ opposed-phase imaging at
3T
Due to the change of the resonance frequency, TEs for in-phase
and opposed-phase conditions at 3T have to be adjusted accordingly.
The values are almost inverted compared with the conditions at
1.5T. In-phase conditions are found at TEs of 2.3 msec and 4.6
msec, while opposed-phase conditions are present at 1.1 msec, 3.5
msec, and 5.75 msec. Therefore, dual-echo sequences are
problematic, since the echo spacing between the initial
opposed-phase echo at 1.1 msec and the following in-phase echo of
2.3 msec are too close for acquisition within a single readout.
Optimization of fat suppression at 3T
There are opposite effects at 3T that improve as well as
deteriorate the quality of fat suppression. The different resonance
frequency at 3T produces a stronger chemical shift between fat and
water, thus facilitating the use of spectral fat suppression. At
the same time,
B
1
homogeneity decreases with
B
0
field strength, which results in a more pronounced variation of the
flip angle over the field of view. This might result in a less
optimal fat suppression, which is particularly problematic for TSE
sequences, while this effect is less problematic with GRE
techniques. In addition, as the field increases, stronger eddy
currents are generated that further reduce the quality of fat
suppression. In daily practice, fat suppression remains
inhomogeneous; nevertheless, this does not lower the accuracy of
imaging evaluation of liver diseases, since the inhomogenous fat
suppression mostly affects the structures of the abdominal wall
only (Figure 2). A recent study the authors conducted showed that,
despite the inhomogeneous fat suppression and other specific
artifacts at 3T, the delineation of liver lesions had no
significant difference between 3T and 1.5T images (Figure 3).
23
Further solutions for optimized fat suppression are improved
gradient-coil design (generating less eddy currents) as well as an
alternative, non-RF-dependent type of fat suppression. This
includes, for example, the Dixon method, which uses several
in-phase and opposed-phase images or inversion recovery-type
techniques.
24
Increased chemical shift at 3T
As mentioned above, the different resonance frequency at 3T
produces a stronger chemical shift between fat and water. This can
be advantageous for the use of spectral fat suppression and for MR
spectroscopy or chemical shift imaging; however, chemical-shift
artifacts are also more pronounced at fat/water interfaces. This
increase of chemical-shift artifacts can be addressed by using a
higher receiver bandwidth, with the consequence of losing some of
the SNR advantages gained by 3T.
Reduction of SAR
As mentioned earlier, the 3 most important techniques for
reduction of SAR are new pulse sequences that apply the variable
flip angle-technique (VFA) or the hyperecho technique (which is a
modification of the former), and parallel imaging (which is
discussed below). The VFA technique was initially described by
Mugler et al
25
and is particularly used for SSFP sequences that offer high SNR
efficiency if large flip angles are applied (α = 60°). However, the
SAR also increases quadratically with the flip angle. Since the
image contrast is mainly affected by the center of k-space, the
flip angle is varied along the phase-encoding direction, ensuring a
large flip angle in the central parts of k-space, while the flip
angle is continuously decreased for the other segments of k-space.
This results in a reduction of SAR by at least 50%. The main
disadvantage of this technique is that SNR is also reduced by
approximately 30%, thus limiting the benefit of the normally
twofold SNR gain at 3T compared with 1.5T. A further refinement of
the VFA technique is the hyperecho technique, which is used for
multiecho sequences such as TSE sequences with long echo trains or
single-shot acquisitions (eg, half-Fourier acquisition single-shot
turbo spin echo [HASTE], rapid acquisition with relaxation
enhancement [RARE]). In this technique, magnetization can be
completely rewound after any arbitrary sequence of RF pulses.
18,19
Only the echoes that encode for the center of k-space are refocused
by 180° RF pulses. Stimulated echoes are also obtained because of a
series of RF pulses with low flip angles. These stimulated echoes
are used for the image reconstruction together with the
conventional echoes as long as they are in-phase with the latter.
This technique results in the reduction of SAR by 60% to 80%, while
the reduction of SNR is virtually negligible. Simpler techniques to
reduce the SAR include the global decrease of the refocusing flip
angle in TSE sequences from 180° to a value between 100° and 120°.
Compared with the hyperecho approach, however, a larger SNR
decrease must be accepted. In fast GRE sequences, the SAR can be
reduced by choosing a longer RF pulse, since the RF pulse power
scales inversely with the square of the pulse length. The primary
parameters used in our institution are T2W fat-saturated TSE with
breath-hold, TR of 3500, TE of 103l, and a flip angle of 140°. For
a respiratory-triggered T2W fat-saturated TSE, we recommend the use
of the following parameters: TR according to the respiratory cycle,
TE of 100, and a flip angle of 140°. Only rarely do these
parameters have to be changed (namely, decrease of the flip angle)
because of SAR limitations.
Reduction of dielectric resonance effects
Dielectric resonance effects are caused by local eddy currents
because of the increased conductivity of the tissue. These effects
are more enhanced at 3T, since the RF waves have a shorter length
as compared with 1.5T. This causes local inhomogeneous areas,
particularly in larger anatomic areas such as the abdomen, where
the wavelength is approximately only half the diameter of the body,
particularly for patients with low fat content. Nevertheless, this
does not necessarily influence the diagnostic value of the
examination, as recent studies with excellent image quality at 3T
and 1.5T have shown. Zech et al
23
stated that despite dielectric artifacts, there has been no
relevant difference in the diagnosis of liver lesions. More
precisely, the delineation of liver lesions in the left lobe, where
the dialectic artifact is frequently present, has been as accurate
as the delineation at 1.5T.
23
In the MR images, dielectric effects are predominantly noticeable
in the ventral aspects of the abdomen as focal areas of varied
signal intensity. They can be effectively reduced by the placement
of an external dielectric gel pad (see above) on the anterior body
surface, which increases the resistance for focal eddy currents.
16
For standard abdominal imaging, this improvement is usually
sufficient; however, there are certain virtually unavoidable
conditions in which substantial artifacts lead to nondiagnostic
image quality. These include physiologic or pathologic conditions
that involve large amounts of intra-abdominal or intrapelvic fluid,
such as pregnancy with a large amount of amniotic fluid or
intra-abdominal ascites. In those cases, images can be almost
nondiagnostic because of focal signal intensity drop-outs.
26
Parallel acquisition techniques
With PAT, the SNR is reduced by the square root of the
acceleration factor times the so-called g-factor, which represents
a measure of the local noise amplification, mostly related to array
coil design characteristics.
4,7,8
Thus with higher acceleration factors of 3, for example, the SNR is
reduced to at least 58% of the original value. This causes a
visible increase of image noise in the images at 1.5T. Because of
the higher SNR at 3T, image noise at comparable acceleration
factors is barely visible in the images. Further development can be
expected from dedicated multichannel MR systems, such as the
Magnetom TRIO with TIM (Siemens Medical Solutions, Erlangen,
Germany and Malvern, PA). An initial study has shown that the new
multielement array (matrix) coils allow for the application of
acceleration factors of up to 4 without noticeable loss of image
quality from increased image noise.
20
With all of these optimization strategies at 3T, a robust image
quality can be obtained in the majority of patients with slice
thickness between 4 and 6 mm, matrix sizes between (320)
2
and (384)
2
, and PAT acceleration factors between 2 and 3. With
multi-breath-hold imaging, the larger abdominal organs, such as the
liver, can be covered in 3 breath-holds, each lasting 11 seconds,
using PAT. For single-shot TSE techniques, such as HASTE (Figure
4), the effective reduction of SAR by the hyperecho technique
allows the acquisition of 15 to 20 slices in a single breath-hold
with a (384)
2
matrix without exceeding the SAR limitations. Slice thicknesses
down to 4 mm are feasible even when PAT is used. The overall
results for high-resolution images with a decreased amount of
blurring is better with the hyperecho technique, as it allows for
the use of higher flip angles compared with the standard VFA
technique with reduced flip angles (down to 60°) for the refocusing
pulse and, therefore, alternated image contrast. For free-breathing
techniques, the use of PAT for high-resolution images with matrix
sizes of (384)
2
allows for total acquisition times of <1.5 minutes.
The combination of several of these described techniques can be
used to optimize dedicated techniques for abdominal imaging, such
as 3D TSE MR cholangiopancreatography (MRCP). For this type of
sequence, a free-breathing acquisition is performed with navigator
correction. To effectively reduce SAR in view of the long echo
trains, a variable flip angle technique is applied. For maintaining
high SNR, the remaining magnetization after the readout is restored
into the longitudinal direction by a dedicated 90° restore pulse.
The SNR gain from those restore-pulses in combination with the
higher field-strengths allows the application of PAT factors of up
to 4. Thus, 3D data sets with an isotropic voxel size of (0.9 mm)
3
can be acquired in only 2 minutes, allowing 3D assessment of the
entire biliary tree by a variety of postprocessing possibilities,
such as multiplanar reformats (MPR), maximum-intensity projections
(MIP), surface-shading techniques (SSD), or volume-rendering
techniques (VRT).
Abdominal and renal MRA requires high spatial resolution
acquired within a short scan time to accurately grade stenoses and
to display subtle vessel wall changes such as fibromuscular
dysplasia.
27
At 3T, high-resolution MRA of the abdominal vessels is feasible
with voxel sizes of 0.9 × >0.8 × 0.9 mm
3
with a PAT acceleration factor of 3 in 18 seconds on a routine
basis.
28
Despite the small voxel sizes (which can be reduced by about 20%
compared with high-resolution MRA at 1.5T
29,30
) and despite the application of a PAT acceleration factor of 3,
sufficient SNR is still present at 3T.
31
Apart from the increased SNR at 3T, MRA also benefits from the
prolonged T1-times of most tissues, while the relaxivities of the
contrast agents are only slightly lowered. Overall, this leads to
an increased contrast agent efficacy at 3T. Preliminary data
suggest that the administered amount of contrast agent can be cut
back without impairing the image quality.
29
In addition, the high SNR at 3T in combination with new acquisition
techniques, such as time-resolved imaging of contrast kinetics
(TRICKS),
32
may foster the widespread use of time-resolved MRA studies at
3T.
Problems for MRA at 3T are twofold. First, the typically short
TRs of <4 msec may result in a high RF power deposition, which
easily exceeds the SAR limits or the stimulation threshold. One
possible approach is to lengthen the RF pulse and the TR and to use
PAT to compensate for the consecutive increase in scan time;
however, this changes the contrast of the image, since a longer
repetition time increases the background signal. This might
decrease the vessel-background contrast and limit the use of MIPs
for the diagnostic evaluation of the MRA. Similarly, decreasing the
flip angle, which may be necessary to stay within the SAR limits,
might also increase the background signal. Fortunately, GRE
sequences used for MRA are not susceptible to dielectric artifacts,
which are often encountered in T2-weighted sequences. Nevertheless,
in the clinical routine, a compromise can usually be found by
carefully adjusting these parameters such that high-resolution
submillimeter MRA is feasible with good image quality.
27,28,31
This is of high importance, since intermediate spatial resolution
in MRA might lead to insufficient diagnostic accuracy, as shown by
a recent publication on MRA at 1.5T
33
(Figures 5 through 7).
Clinical value of 3T for abdominal MRI
In general, those applications for abdominal MRI seem to profit
from the higher field-strength that critically depends on spatial
resolution. One promising application appears to be the assessment
of multicystic abdominal masses, in which the delineation of
enhancing septa is important for the detection of malignancy.
34
This includes multicystic masses of the kidney and ovaries, in
which small nodular thickening of the septa with increased
enhancement after gadolinium administration is indicative of early
malignancy.
35
With an optimal spatial resolution, the differentiation between
cysts and renal cell carcinoma in von Hippel-Lindau disease, the
differentiation between cystadenoma and cystadenocarcinoma in
cystic ovarian lesions, and the early detection of renal cell
carcinoma in acquired polycystic disease can be facilitated.
34-36
For the assessment of focal liver le-sions, there is still no
consensus whether the high resolution truly converts into a better
detection of small lesions (<1 cm) (Figure 8). However, it has
already been shown in small, noncontrolled studies that a high
spatial resolution (eg, with 3D GRE sequences) helps to improve the
detection of small lesions and the characterization of lesions in
terms of better delineation of morphological features, such as the
vascular architecture or a central scar
37,38
(Figure 9). As high-field MRI (with its high spatial resolution) is
not limited by a decreased liver-to-lesion contrast (as is the case
with computed tomography), it can be expected that the high
resolution should improve overall lesion detection.
39,40
The dilemma at this point, however, is that because of the reduced
availability of 3T in the clinical routine, no larger systematic
studies comparing 1.5 and 3T correlated with an invasive gold
standard, such as biopsy or surgery, have been performed.
An area that holds great promise is the staging of malignant
lesions of the uterine cervix and the rectum.
41
MRI at 1.5T has been notoriously weak in differentiation between
stage IIA cervical carcinoma and stage IB, with an accuracy of only
approximately 80%.
42
This question is highly relevant, since differentiation between
those 2 stages affects the decision for either radical hysterectomy
or primary radiation therapy. Since the extension of the tumor
beyond the cervical stroma, which is sometimes <1 mm in width,
is the discriminator between those 2 stages, one might expect an
improved staging with matrix sizes higher than (512)
2
. Similar improved performance could be expected for staging rectal
carcinoma, in which MRI is known to overstage lesions of TNM stage
T1 as T2 lesions, since reliable differentiation between the
submucosa and muscularis propria layers is difficult at the current
maximum matrix size of (512)
2
.
43,44
Recent publications also re-port that increased spatial resolution
is helpful for the evaluation of the nodal staging in colorectal
cancer because the better visualization of morphologic features
help distinguish malignant from benign lymph nodes.
45
One of the most promising applications is high-resolution 3D
MRCP for diseases of the biliary tree, which predominantly manifest
in the high-order branches in their early stages, such as primary
sclerosing cholangitis (PSC).
46
Reliable visualization of the biliary tree with MRCP at 1.5T is
possible only up to the second-order branches with current
techniques.
47
Compared with endoscopic retrograde cholangiopancreatography
(ERCP), MRCP has been proven to be cost-effective, especially
because of its lack of any side effects and complete
noninvasiveness. Despite this, it has not established a role for
the initial detection of PSC disease manifestation.
48
This patient group, however, particularly benefits from the
noninvasive techniques, since they are frequently young patients
and early treatment can substantially affect the cause of their
disease.
Untreated patients may experience severe complications, such as
multiple strictures of the intra- and extrahepatic bile ducts,
which ultimately may progress to cirrhosis and liver failure.
49
Moreover, in these patients, ERCP for initial diagnosis can lead to
substantial side effects such as pancreatitis or common bile duct
perforation, which is observed more frequently than in other
patients undergoing ERCP.
50
With voxel sizes of >0.9 × 0.9 × 0.9 mm
3
spatial resolution and images free of motion artifacts with scan
times of roughly 2 minutes, 3D MRCP at 3T now allows for the
detection of high-order branches of the biliary tree in volunteers
without biliary disease. Our own preliminary data show that focal
dilatations and stenoses of third- and fourth-order branches, which
are often the early manifestation of PSC, can be well detected with
good correlation to ERCP. The acceleration of the acquisition by a
PAT factor up to 4 makes this an easily feasible technique for
these patients. MR cholangiopancreatography can also be combined
with high-resolution liver imaging at 3T with a slice thickness of
only 3 mm, which may allow for the earlier detection of malignant
tumors of the bile duct, such as cholangiocarcinoma.
Ultra-high-resolution contrast-enhanced breath-hold MRA with
0.7- to 0.8-mm isotropic resolution might eliminate the need for
diagnostic digital subtraction angiography (DSA) of the mesenteric
and renal arteries. A recent publication in the
Annals of Internal Medicine
has shown that MRA at 1.5T with a current spatial resolution of
only 1.5 mm or even worse is not sufficient to accurately diagnose
and stage the degree of renal artery stenosis.
33
Initial results at 1.5T with submillimeter (0.8 × 0.8 × 0.9 mm
3
) isotropic spatial resolution and reformats of the vessel area
perpendicular to the vessel axis show that assessment of the area
stenosis is a much more reliable parameter for exact grading of the
degree of stenosis than for grading the diameter stenosis in-plane.
14
It is expected that this improvement will further increase at
3T.
Conclusion
One has to accept that imaging of the abdomen at 1.5T on a
state-of-the-art system with parallel imaging, navigator
correction, and an advanced sequence design is a tough competitor
for the evolving field of 3T. Nevertheless, it can be stated that
it is now feasible to acquire single-shot TSE images with PAT and
hyperecho technique at a very high resolution with minimal blurring
and within the SAR limits. Those images appear to be superior to
those currently acquired at 1.5T (as shown in Figures 8 and 9).
Parallel acquisition techniques are generally applicable without
major SNR constraints, allowing for the use of higher acceleration
factors for shorter breath-holds. The fat suppression appears
acceptable despite problematic dielectric resonance effects. For
techniques such as 3D TSE MRCP, ultra-high-resolution submillimeter
studies with minimal acquisition times are feasible and may also
eliminate the use of diagnostic ERCP for diseases, such as primary
sclerosing cholangitis, in the future. No larger systematic studies
have yet been performed for the detection and characterization of
small focal liver lesions. It appears, however, that there is
improvement at 3T, since in MRI the lesion delineation is primarily
limited by the spatial resolution rather than by the
liver-to-lesion contrast because of the excellent soft-tissue
contrast. The resolution of contrast-enhanced 3D MRA approximates
that of DSA for the first time, which should eliminate the use of
diagnostic renal or mesenteric DSA other than for rare diseases,
such as small vessel vasculitis. In the pelvis,
ultra-high-resolution imaging with SE and TSE sequences might
further improve the staging of early T1 and T2 tumors of the
cervix, uterus, and rectum. The true gain, however, in clinical
value is still speculative at the present, since no large
comparative studies have yet been performed or published.