Dr. Balci
is an Assistant Professor, Department of Radiology, Saint Louis
University School of Medicine, St. Louis, MO.
Dr. Semelka
is a Professor and
Dr. Altun
s a Research Fellow, Department of Radiology, The University of
North Carolina at Chapel Hill, NC.
Image quality, reproducibility of image quality, and good
conspicuity of disease requires the use of sequences that are
robust and reliable and that avoid artifacts.
1-5
Maximizing these principles to achieve high-quality diagnostic
magnetic resonance (MR) images usually requires the use of fast
scanning techniques with diagnostic image quality. Respiration,
bowel peristalsis, and vascular pulsations result in artifacts that
have lessened the reproducibility of MR imaging (MRI).
Breathing-independent sequences and breath-hold sequences form the
foundation of high-quality MRI studies of the abdomen.
Disease conspicuity depends on the principle of maximizing the
difference in signal intensities between diseased tissues and the
background tissue. For disease processes situated within or
adjacent to fat, this is readily performed by manipulating the
signal intensity of fat, which can range from low to high on both
T1-weighted (T1W) and T2-weighted (T2W) images. For example,
diseases that are moderate to high in signal intensity on T2W
images, such as inflammatory exudates or ascites, are most
conspicuous on sequences in which fat signal intensity is rendered
low with fat suppression.
Gadolinium chelate enhancement may routinely be useful because
it provides at least 2 further imaging properties that facilitate
the detection and characterization of disease-specifically, the
pattern of blood delivery (ie, capillary enhancement) and the size
and/or rapidity of drainage of the interstitial space (ie,
interstitial enhancement).
6
Capillary-phase image acquisition is achieved by using a
short-duration sequence initiated immediately after gadolinium
injection. The majority of focal mass lesions are best evaluated in
the capillary phase of enhancement, particularly lesions that do
not distort the margins of the organs in which they are located
(eg, focal liver, spleen, or pancreatic lesions). Images that are
acquired at 1.5 to 10 minutes after contrast is administrated are
in the interstitial phase of enhancement; the optimal window is 2
to 5 minutes postcontrast. Diseases that are superficial,
spreading, or inflammatory in nature are generally well shown on
interstitial-phase images. The concomitant use of fat suppression
serves to increase the conspicuity of disease processes that are
characterized by increased enhancement on interstitial-phase
images, including peritoneal metastases, cholangiocarcinoma,
ascending cholangitis, inflammatory bowel disease, and
abscesses.
MRI strategies are also very important to obtain sufficiently
diagnostic images from all patients, including noncooperative ones.
The great majority of diseases can be characterized by defining
their appearance on T1W, T2W, and early and late postgadolinium
images. In this way, the variety of information describing the
different features of diseases and organ systems can be obtained
comprehensively. Also, with the use of a diverse group of sequences
acquired in multiple planes, target lesions can be characterized
and the likelihood of not detecting or of misclassifying disease is
minimized. The length of examination time should be short to
increase patient cooperation and throughput; therefore,
shorter-duration breath-hold or breathing-independent sequences
should be applied in the examination protocols. To decrease the
length of examinations and to make them more reproducible and
standard, the use of set protocols is needed. In addition, the
protocols should have some minor redundancy to allow for the
possibility that some of the sequences may be nondiagnostic
(because of some patient factors, such as motion), and so the
remaining sequences should provide sufficient diagnostic
information. The MR examination protocols differ for the upper
abdomen and the pelvis as well as for the disease processes
involved. When only one phased-array coil is used for an
examination that is to include the upper abdomen and pelvis, the
most efficient way to perform the examination is to complete the
upper abdomen study first and continue with the pelvis later. In
this way, table motion and repositioning of the coil are minimized.
If multiple phased-array coils are used simultaneously together
with remote table motion, as in current MRI systems, there will be
no need for repositioning the coil, so the length of examination
will be decreased and precontrast pelvis images may be obtained, if
necessary (eg, in the evaluation of endometriosis). With the use of
phased-array multicoils, both slice thickness and field of view can
be substantially modified for many patients.
One of the most important recent developments is the parallel
imaging technique, which requires specialized multichannel coils.
The benefits of decreasing sequence duration and increasing spatial
resolution are only now beginning to be realized. Dramatic
improvements in these areas are expected in the near future and
will likely greatly increase the role of body MRI.
The basic approach to all body MRI examinations is to obtain at
least one noncontrast T1W sequence (a standard non- fat-suppressed
sequence is always advisable; often, out-of-phase and
fat-suppressed sequences are useful additions), one T2W sequence
(often a single-shot echo-train technique is sufficient), and
immediate postgadolinium and 2-minute postgadolinium T1W
gradient-echo sequences. A brief description of these sequences
follows.
T1-weighted sequences
T1-weighted sequences are routinely useful for investigating
diseases of the liver. The primary information that precontrast T1W
images provide includes: 1) information on abnormally increased
fluid content or fibrous tissue content that appears low in signal
intensity on T1W images; and 2) information on the presence of
subacute blood or concentrated protein, which are both high in
signal intensity. T1-weighted sequences obtained without fat
suppression also show the presence of fat as high-signal-intensity
tissue. The routine use of an additional fat-attenuating technique
facilitates reliable characterization of fatty lesions.
Spoiled gradient-echo sequences
Spoiled gradient-echo (SGE) sequences are the most important and
versatile sequences for studying liver disease. Image parameters
for SGE are: 1) relatively long repetition time (TR) (approximately
150 msec) to maximize both the signal-to-noise ratio and the number
of sections that can be acquired in one multisection acquisition;
and 2) the shortest in-phase echo time (TE) (approximately 6.0 msec
at 1.0T and 4.2 to 4.5 msec at 1.5T) to maximize both the
signal-to-noise ratio and the number of sections per acquisition.
2
Hydrogen protons in a voxel that contains 100% fat will precess
approximately 220 to 230 Hz slower than a voxel composed of 100%
water, at 1.5T. That means that every 4.4 msec, the fat protons
will lag behind by 360˚and will regain in-phase orientation
relative to the water protons, while at 2.2 msec, or at half this
time, the fat and water protons will be 180˚out-of-phase.
Current-generation MR software programs have incorporated dual-echo
breath-hold SGE sequences that can acquire 2 sets of k-space filled
to obtain 2 sets of images-one set in-phase, the other
out-of-phase-with spatially matched slices. For routine T1W images,
in-phase TE may be preferable to the shorter out-of-phase TEs (4.0
msec at 1.0T and 2.2 to 2.4 msec at 1.5T), to avoid both
phase-cancellation artifact around the borders of organs and
fat-water phase cancellation in tissues containing both fat and
water protons. The flip angle should be approximately 70˚to 90˚, to
maximize T1W signal. With the use of the larger built-in body coil,
the signal-to-noise ratio of SGE sequences is usually suboptimal
with a section thickness <8 mm, whereas with the phased-array
surface coils, a section thickness of 5 mm results in
diagnostically adequate images.
Out-of-phase (opposed-phase) SGE images are useful for revealing
diseased tissue in which mixtures of fat and water protons are
present within the same voxel. A TE of 2.2 msec is advisable at
1.5T, and 4.4 msec is advisable at 1.0T. A TE of 6.6 msec is also
out-of-phase at 1.5T, but the shorter TE of 2 msec is preferable
because of decreased susceptibility effects. In combination with a
T2W sequence, it is easier to distinguish fat and iron in the
liver. The most common indications for out-of-phase imaging are the
detection of abnormal fat accumulation within the liver and the
detection of lipid within adrenal masses, a feature used to
characterize benign adrenal adenomas (Figure 1).
Fat-suppressed SGE sequences
Fat-suppressed SGE sequences are routinely used as precontrast
images for evaluating the pancreas and for the detecting subacute
blood (Figure 2). Image parameters are similar to those used for
standard SGE. It may be advantageous to employ a lower out-of-phase
TE (2.2 to
2.5 msec at 1.5T), which benefits from additional
fat-attenuating effects and also increases signal-to-noise ratio
and the number of sections per acquisition.
4
On current MRI machines, fat-suppressed SGE may acquire 22 sections
in a 20-second breath-hold, with reproducible uniform fat
suppression. One method that modern systems use to reduce the
amount of additional time that fat suppression adds to the SGE
sequence and to acquire a greater number of slices per breath-hold
is to perform a fat-suppression step only after several
phase-encoding steps, instead of after every phase encode. Another
approach is to selectively tune the stimulation radiofrequency (RF)
pulse to activate only protons in water, but not in fat, thus
eliminating the need to add fat-saturation pulses. This technique
is referred to as
water excitation
.
Fat-suppressed SGE images are used to improve the contrast
between intra-abdominal fat and diseased tissues and blood vessels
on interstitial-phase gadolinium-enhanced images. Gadolinium
enhancement generally increases the signal intensity of blood
vessels and disease tissue, and fat suppression diminishes the
competing high signal intensity of background fat.
3-dimensional gradient echo
Three-dimensional (3D) gradient echo (GE) imaging has been used
extensively for MR angiography (MRA), but only recently has evolved
into an accepted, useful technique for imaging the liver. This
development has partly been achieved simply by reducing the flip
angle from 70˚to 90˚, down to 12˚ to 15˚. Advantages include the
ability to acquire a volumetric data set that can be sectioned into
thinner sections than are typically used for 2-dimensional (2D)
images, generally in the 2.5 to 3.0-mm-per-slice range, with
contiguous slices, and with images that can be postprocessed into
other imaging planes (Figure 3). Although there are differences
between some of the sequence features seen between various MR
systems, fat suppression tends to be superior with greater
uniformity, as compared with 2D SGE. On some MR systems, it is also
possible to image a larger volume of tissue during the same
breath-hold period than is possible with 2D SGE. A potential
limitation of 3D GE imaging is a diminished contrast-to-noise
ratio. This has led to concern regarding the use of this technique
other than for gadolinium-enhanced fat-suppressed
interstitial-phase imaging, in which the gadolinium effectively
improves the contrast-to-noise ratio.
5-8
Motion-insensitive GE
One limitation of GE sequences, both 2D and 3D, is relative
motion sensitivity, and, thus, the requirement for cooperation by
the patient in following breathing instructions. In uncooperative
patients, GE sequences may be modified as a single-shot technique
using the minimum TR to achieve breathing-independent images
(Figure 4). Such sequences have included so-called
magnetization-prepared rapid-acquisition gradient echo (MP-RAGE),
and turbo fast low-angle shot (turboFLASH). This technique has been
achieved using magnetization-prepared GE, in which an inversion
prepulse leads to the ability to improve T1W contrast during a
short single-slice acquisition. As the protons recover
magnetization, a single-slice short TR GE imaging sequence is
performed. An inversion time of approximately 0.5 second provides
optimal T1W contrast and sufficient time to allow the protons to
recover between slices, which leads to an effective slice-to-slice
TR of ≥1.5 seconds. This technique can be performed to yield either
bright or dark through-plane flowing blood, by making the prepulse
either slice-selective or nonselective, respectively. Limitations
of this technique have included the inability to obtain as high or
as predictable T1W contrast as with standard SGE. Another
limitation is that the magnetization-prepared GE slice-by-slice
technique is not ideal for dynamic gadolinium-enhanced imaging of
the liver, particularly during the hepatic-arterial dominant phase.
As each slice requires approximately 1.5 seconds for acquisition,
the time difference accumulated between the top and bottom liver
slices is too great to capture the entire liver in the arterial
phase of enhancement.
4
Nonetheless, slice-selective inversion-pulse single-shot T1W GE
allows contiguous acquisition of slices without a need for time
delay between slices.
T2-weighted sequences
The predominant data provided by T2W sequences are: 1) the
presence of increased fluid in diseased tissue, which results in
high signal intensity; 2) the presence of chronic fibrotic tissue,
which results in low signal intensity; and 3) the presence of iron
deposition, which results in very low signal intensity.
Echo-train spin-echo sequences
The principle of echo-train spin-echo sequences is to summate
multiple echoes within the same TR interval to decrease examination
time, increase spatial resolution, or both. We routinely use a
single-shot technique for liver imaging: either half-Fourier
acquisition single-shot turbo spin echo (HASTE) or single-shot fast
spin echo (SSFSE). This is a slice-by-slice technique in which a
single slice-selective excitation pulse is followed by a series of
echoes, typically using between 80 and 100 180˚pulses, each
separated by approximately 3 msec, to fill in k-space for the
entire slice. The T2W contrast is achieved by using the echoes
obtained approximately 80 to 90 msec for filling central k-space,
where central k-space is responsible for image contrast. Although
the theoretical TR is infinite, each slice requires roughly 1.2 to
1.5 seconds before continuing to the next slice. However, the
motion-sensitive component represents only a smaller fraction of
the entire acquisition period, which makes this technique
relatively insensitive to breathing or other motion artifacts. The
multiple 180° pulses serve to minimize magnetic susceptibility by
limiting T2* decay. Echo-train spin-echo has achieved widespread
use because of these advantages.
In contrast, conventional T2W spin-echo sequences are lengthy
and suffer from patient motion and increased examination time. The
major disadvantage of echo-train sequences is that T2 differences
between tissues are decreased. In the liver, the T2W signal
difference between diseased and background normal liver may be
small, and the T2-averaging effects of summated multiple echoes
blur this T2 difference. This results in relatively diminished
lesion conspicuity for lesions with mildly elevated T2 signal
intensity, such as hepatocellular carcinoma, as compared with
standard spin-echo sequences. Fortunately, diseases with T2 values
similar to those of liver generally have longer T1 values than
liver, so that lesions poorly visualized on echo-train spin echo
are generally well visualized on SGE or immediate postgadolinium
SGE images as low-signal lesions.
Echo-train spin-echo sequences-and T2-weighted sequences, in
general-are important for evaluating the liver. In liver masses,
T2W images are predominantly important for lesion characterization,
while T1W images are important for both lesion detection and
characterization. T2weighted images also are important for the
assessment of diffuse liver disease, including iron deposition,
edema related to active liver disease, and fibrosis. Echo-train T2W
sequences are important for assessment of fluid-filled structures,
including the bile duct, gallbladder, and pancreatic duct, and the
stomach and bowel as well as cysts or cystic masses, abscesses or
collections, or free fluid in the abdomen or pelvis (Figure 5). The
relative resistance of echo-train images to motion degradation
generally yields better resolution of structures internal to cystic
masses, such as the septations within a pancreatic serous or
mucinous tumor. MR cholangiopancreatography is based on modified
echo-train sequences, in which the effective TE becomes longer-
approximately 250 to 500 msec. Lengthening the TE results in
heavily T2W high-contrast images that yield most soft tissues dark
and that make fluid in the bile ducts, gallbladder, and pancreatic
duct very bright. MR cholangiopancreatography can be performed in
thin sections (3 to 4 mm) for higher resolution, or using a single
thick slab of 3 to 4 cm, to include the majority of the pancreatic
and bile ducts in a single image. T2-weighted single-shot
echo-train imaging is well suited to studying the bowel because of
insensitivity to both respiratory motion and bowel peristalsis, and
the relative resistance to the distorting paramagnetic effects of
intraluminal bowel gas because of the repeated refocusing echo
pulses.
Fat is high in signal intensity on T2W echo-train spin-echo
sequences when compared with conventional spin-echo sequences, in
which fat is intermediate in signal intensity. Fat may also be
problematic in the liver because fatty liver will be high in signal
intensity on echo-train spin-echo sequences, thereby diminishing
contrast with the majority of liver lesions, which are also
generally high in signal intensity on T2W images. It may be
essential to use fat suppression on T2 echo-train spin-echo
sequences for liver imaging. Fat suppression should generally be
applied to at least 1 set of images of the liver to ensure optimal
contrast between high-signal abnormalities (such as fluid
collections or cystic masses) and adjacent intra-abdominal or
pelvic fat.
3
Conclusion
MRI performed using a combination of various short-duration
sequences employed in a set imaging protocol will detect the great
majority of disease processes in the abdomen and pelvis with very
high sensitivity and specificity and with high patient safety. With
the use of motion-insensi-tive T1W and T2W sequences, MRI has a
very high diagnostic accuracy even in noncooperative patients. The
development of volumetric imaging sequences together with the
development of faster and artifact-insensitive sequences employing
parallel imaging will even further increase the diagnostic role of
MRI in the near future.