Dr. Crawley
is an Assistant Professor,
Mr. Poublanc
is a Research Assistant,
Mr. Ferrari
is a Research Assistant, and
Dr. Roberts
is an Associate Professor in the Department of Medical Imaging,
University of Toronto and University Health Network, Toronto,
Ontario, Canada.
Perhaps no techniques better exemplify the successful transition
from laboratory experiment to clinical routine than diffusion and
perfusion-sensitive magnetic resonance imaging (MRI). Little more
than a decade ago, these techniques were the exclusive province of
esoteric physics. Yet, they have emerged as an invaluable asset to
the neuroradiologist and neurologist alike; in some sense they can
be considered to be at the forefront of the movement toward
physiological, or functional, imaging. The purpose of this article
is to review the physical basis of the techniques, to discuss their
implementation and future potential, and to consider their
principal application in the diagnosis and, importantly,
characterization of acute cerebral ischemia.
Diffusion and apparent diffusion
Like all fluids, water molecules undergo continual random motion
(so-called Brownian motion) at a rate described by the
self-diffusion coefficient, D, and the (lesser-known) Einstein
equation (<L
2
> = 2D*). This motion arises largely because of the thermal
energy possessed by the molecules being expended as kinetic energy
and, thus, creating movement.
1
If the position of a particular molecule is known at a given time,
the Einstein equation predicts the expectation value of the
molecule's displacement (L) at a later time (*), based on the value
of the self-diffusion coefficient, D. Using paired magnetic field
gradient pulses to effectively encode, and then subsequently
decode, spatial position, the degree of water displacement, or
diffusion, can be estimated in MRI, via the relative
ineffectiveness of spin-echo formation (and thus the signal loss on
such a diffusion-weighted image [DWI])
2-5
(Figure 1).
Formally, the signal intensity observed on a DWI is approximated
by the expression S * e
-bD
, where the factor b relates to the degree of diffusion sensitivity
of the sequence.
6
Consider a T2-weighted MRI--the value of echo time (TE) effectively
determines the sensitivity of the sequence to the process of
T2-relaxation; analogously the b-value determines the sensitivity
of a DWI to the process of diffusion: sequences with low b-values
are relatively insensitive to diffusion; sequences with high
b-values are sensitive to even minor water displacements. Typically
encountered bvalues in clinical practice are of the order of 1000
sec/mm
2
. From observing the signal obtained with different b-values
(typically b = 0 and b = 1000 sec/mm
2
), it is thus possible to estimate D. However, if it were truly
water self-diffusion alone, this would be of rather limited
biomedical interestthe self-diffusion coefficient of a molecule
depends on its molecular weight, which is constant for water (18)
and the absolute temperature, T, which is also of rather limited
variation in vivo (310 K, 37°C). In fact, however, the
physicochemical microenvironment of tissue prohibits free
self-diffusion. Consequently, the parameter we derive from the
above analysis is referred to as the apparent diffusion coefficient
(ADC), which is related to microviscosity, organelle, membrane and
molecular interactions, active transport mechanisms, perfusion,
bulk flow, and, of course, the ensemble averaging of the many
different water environments contained within an imaging voxel.
7
By its nature, ADC is also extremely sensitive to gross movement of
the patient. Nonetheless, if gross motion is minimized (by physical
restraint and ultra-fast imaging), derivation of the ADC yields
tantalizing insight into water molecule displacements on a micron
(µm) scale, several orders of magnitude smaller, in fact, than the
nominal pixel, or spatial, resolution of the image.
Anisotropy
Furthermore, while self-diffusion is an isotropic process (ie,
there is an equivalent likelihood of molecular displacement in any
given direction), apparent diffusion may well exhibit a directional
preference (consider, for example, a cylindrical microstructure in
which motion is permitted along the long axis, but impeded in the
cross-sectional plane). Such directional preference is termed
anisotropy and is a characteristic feature, for example, of white
matter tracts in the brain. Anisotropy can be revealed via
diffusion-weighted MRI by conducting separate experiments in which
diffusion sensitivity or positional encoding/ decoding is applied
in successive experiments in different directions (eg, by first
using the x-axis gradients to encode displacement in the
x-direction, then subsequently the y-axis gradients, and so on).
7-9
In fact, the degree of anisotropy and its preferred direction (an
example used by Prof. Jim Provenzale, Duke University Medical
Center, is to consider the shape of an American football versus a
soccer ball) can best be estimated by performing such directional
displacement encoding in multiple (>= 6) directions and
describing the diffusion process as a 3 * 3 matrix, or tensor
10
(Figure 2). Since this analysis can be performed on a
voxel-by-voxel basis it is, thus, possible to ascertain for each
image voxel both the degree of anisotropy (how elongated is the
football) and its preferred direction (which way is it pointing).
Recent sophisticated postprocessing techniques have used these
properties to follow the arrows and essentially reconstruct the
fibers of white matter tracts.
11
Diffusion in stroke
However, the predominant clinical indication for
diffusion-weighted MRI remains the diagnosis of acute cerebral
ischemia.
12,13
While the pathophysiologic mechanisms underlying the diffusion
changes remain the subject of debate, a simplistic description of
the process is offered below: Arterial occlusion or sustained
severe stenosis leads to hypoperfusion of the subserved vascular
territory. Subject to such deprivation of essential nutrients,
energy-requiring cell membrane transporters (particularly the Na
+
/K
+
-ATPase pump) begin to fail. The cells thus lose the ability to
regulate their volume and there is an influx of extracellular fluid
into the intracellular spacethe cells swell, a condition referred
to as cytotoxic edema. Whether diffusion is reduced because of
reduced fraction (and increased tortuosity) of the relatively free
diffusive environment of the extracellular space, or because of the
increased volume fraction of the slower diffusive environment of
the intracellular space, or because of reduced activity of membrane
transporters due to energy deprivation, or all of the above, the
net consequence is that the ADC may be reduced by up to 50% within
minutes of the insult.
14,15
As a consequence of reduced freedom of diffusion in this ischemic
territory, the region appears characteristically relatively
hyperintense in comparison with healthy tissue on
diffusion-weighted MRI (Figure 3). It is worth noting that a DWI
formed by the application of pulsed gradients to a conventional
spin-echo or spin-echo echoplanar imaging (SE-EPI) sequence incurs
both diffusion and T2-sensitivity. Consequently, hyperintensity
apparent on a DWI could, in principle, be attributed either to
reduced diffusion, or simply to elevated T2 (so called T2-shine
through). Consideration of a T2-weighted (but not
diffusion-weighted) image (namely the control image acquired with b
= 0) reveals the origin of the hyperintensity and construction of
the synthesized ADC map (derived from two or more images with
different b-values, typically b = 0 and b = 1000 sec/mm
2
) quantitatively eliminates any influence of T2-weighting.
16
In the absence of adequate intervention, the pathophysiologic
cascade continues and cell lysis ultimately occurs, leading to the
observation that the area of reduced diffusion visualized by DWI
predicts the area of ultimate infarct. So, while DWI is sensitive
to the tissue or cellular consequences of hypoperfusion, the root
of the problem lies in the effective delivery of blood to the
tissue. Consequently, DWI is commonly accompanied by techniques
that attempt to visualize and characterize perfusion itself.
Perfusion
From a physiological point of view, perfusion is an intuitively
easy concept to grasp; for any volume of tissue under
consideration, a measure of perfusion should estimate the volume of
blood that passes through the capillary bed per unit time. It is
this volume of blood that delivers nutrients to the tissue, whereas
blood that travels straight through the volume of tissue within
arteries or veins should not be included in the measurement. Thus,
importantly, perfusion is not exactly the same as blood flow. For
quantitative work in animal models, radioactively labeled
microspheres are injected into the blood in order to mimic this
delivery mechanism to the tissue. The diameter of the micro-spheres
is selected so that they become trapped within the capillaries and
the level of radioactivity is subsequently measured for each tissue
of interest. Any noninvasive MRI method should aim to achieve a
similarly meaningful estimation of perfusion.
We have already shown how MRI can be made sufficiently sensitive
to the random motion of water molecules across an applied
magnetic-field gradient to enable the apparent diffusion
coefficient of water within the tissues to be imaged. As a first
step in directing the motion-sensitivity of the MRI signal
specifically toward flow through the capillaries, gradient pulses
with a relatively small sensitivity to diffusion (rather low
b-value) were applied in order to measure the component of apparent
diffusion of flow through the quasi-random tortuosities of the
capillary bed.
17
In practice, a large number of scans over a range of b-values must
be acquired in order to separate out this apparent diffusion
component from the underlying smaller diffusion of the water
molecules. Ultimately, the method is limited by the fact that the
proportion of water within the capillary bed is only approximately
2%, making its contribution to overall apparent diffusion extremely
hard to quantitate reliably above the noise in the images.
In a quite different approach, the problem of sensitivity has
been solved by employing an intravenous (IV) bolus injection of
gadolinium-based contrast agent that profoundly affects the signal
(by up to 50%) on a gradient-echo (typically gradient-echo
echoplanar imaging [GE-EPI]) scan as it passes through the
vasculature.
18,19
A power injector may be used to achieve a very tight bolus, so that
the signal loss during the first pass through the tissue can be
measured relative to an initial baseline by a high temporal
resolution (approximately 1 to 2 sec) scan (repeated successively
for a total scan time of approximately 1 min). Gadolinium is a
paramagnetic substance, and in the chelated form in which it is
employed as a contrast agent, it does not cross the normally intact
blood-brain barrier (BBB) and, hence, remains intravascular within
the brain. As the bolus passes through the blood vessels, a
relatively large magnetic field difference is created between the
gadolinium-doped blood and the surrounding brain tissue. The
different magnetic fields within each voxel cause the MRI signal to
dephase, producing a progressive signal loss (characterized by a
time constant T2*) over the TE. A gradient-echo scan with long
echo-time (approximately 40 to 60 msec) is referred to as a
T2*-weighted scan and is the basis of the dynamic susceptibility
contrast (DSC) approach to perfusion sensitive MRI
20
(Figure 4).
This kind of scan is extremely sensitive to the local field
inhomogeneities produced by the contrast agent, but this signal
loss mechanism is not particularly specific to capillary vessels.
Over the last several years, it has become somewhat more common to
opt for a T2-weighted spin-echo scan, which produces a signal loss
that mostly arises from the passage of gadolinium through the
capillaries rather than through the larger blood vessels within the
voxel.
21
Since the spin-echo refocuses static field inhomogeneity effects,
the only remaining signal loss mechanism is that due to water
diffusion through the susceptibility gradients. Since there is a
limited time window (ie, the TE) for this process to occur, the
spin-echo scan will be sensitive only to susceptibility gradients
that are significant over the short distance that a water molecule
diffuses during the TE. This corresponds fairly closely to the
susceptibility effect around a gadolinium-containing capillary.
Nonetheless, this capillary specificity is obtained at the expense
of sensitivity, and typically SE-EPI signal losses are considerably
smaller than those observed using a GE-EPI sequence, prompting the
use of higher doses (eg, 0.2 to 0.3 mmol [Gd]/kg) of contrast
agent.
Since the contrast agent remains within the vascular space, it
is not surprising that this dynamic bolus method measures blood
volume more directly than it can measure perfusion. In fact, for a
given quantity of contrast agent passing into the region of
interest, the relative blood volume is basically given by the area
under the curve representing signal loss as a function of time
(after rescaling of signal intensity to concentration
proportionality).
19
This is intuitively obvious: if the bolus is more spread out in
time, the signal loss curve will also be more spread out over time
but the peak signal loss will be correspondingly less, and the area
under the curve will remain the same. Clearly, as long as the blood
vessels are intact and are not totally collapsed, one may refer to
a histologically meaningful blood volume even when there is no
actual blood flow into the tissue. Therefore, while the measurement
of blood volume is useful in terms of issues such as vascular
resistance, on its own it says nothing about whether blood is
actually reaching the tissue.
If the gadolinium could be injected instantaneously into the
arterioles supplying the capillaries, then the temporal behavior of
the tracer concentration curve (specifically the mean transit time
[MTT]) would provide the necessary information for a calculation of
the volume flow rate (ie, perfusion) through the capillaries.
22
Specifically, the central volume principle states that perfusion
(volume of blood/time) = blood volume/MTT.
23
Unfortunately, the IV bolus requires several seconds of injection
time (typically 2 to 5 sec) and is then dispersed somewhat as it
travels through the heart and pulmonary circulation, so that its
breadth may be as much as 10 sec prior to entering the tissue of
interest. This arterial input function (AIF) can be measured from
the tracer concentration curve within an appropriate artery inside
the imaging volume. A mathematical deconvolution is then required
to attempt to remove the temporal blurring caused by the prior
dispersal of the bolus.
24
Unfortunately, the MTT through each voxel is short compared with
the prior blurring (dispersion) of the bolus, so it is hard to
achieve accuracy in the measurement of either the MTT itself or
thus in the derived estimate of perfusion.
In spite of the fact that a quantitative measure of perfusion is
hard to achieve with the dynamic bolus method, it is widely used in
clinical practice because it offers rather robust qualitative and
indeed semi-quantitative indicators of perfusion.
25,26
We noted above that the direct measurement of blood volume may be
rather irrelevant to the evaluation of a tissue's perfusion status.
Referring again to a situation in which there is some finite blood
volume but little actual movement of blood through the tissue, it
is clear that the dynamic bolus scan will register little or no
signal loss within this tissue, because the bolus will barely
arrive at or pass through its capillaries. For the general case of
reduced perfusion through ischemic tissue, the whole time-scale of
the arrival and local dispersal of the bolus will be spread out. It
will then become more difficult to detect any measurable area under
the tracer concentration curve even if the bolus completes its
passage through the tissue within the duration of the scan, since a
relatively small signal loss occurring over a long time is much
harder to distinguish from other causes of temporal drift in the
MRI signal. Therefore, in the presence of local ischemia, the area
under the tracer concentration curve provides a measure of blood
volume that becomes heavily weighted by some perfusion factor when
perfusion to the tissue is severely compromised.
This still leaves a lack of sensitivity for mild to moderate
perfusion deficits, where the dynamic bolus method is able to
measure the blood volume fairly accurately. In this case there is
little or no perfusion-weighting in the final blood volume image,
which on its own may give little indication of any compromised
blood flow into the tissue. We expect the tracer concentration
curve to be delayed and dispersed in time relative to normally
perfused tissue in the same subject. This should provide some
useful information without explicitly requiring any deconvolution
procedure. We note here that the purpose of the deconvolution
method described above is to attempt to provide absolute
quantitation of perfusion that would enable a measurement from one
scan to be compared with measurements from other scans (obtained
with contrast boli with different temporal profiles), or for the
establishment and use of quantitative cerebral blood flow (CBF)
thresholds for stratifying ischemia. In most clinical situations,
the basic function of a diagnostic imaging test is to map out
regions of pathology relative to more normal areas within the same
individual. While it would be ideal to be able to compare
measurements between scans, in practice, a more robust method can
often be employed if one limits the scope of quantitation to
relative measurements between tissues obtained within an individual
scan. This is certainly the case with the dynamic bolus method,
since for a relative measure of perfusion, the MTT can be
calculated from the tracer concentration curve without the need for
deconvolution, at least assuming all tissues within the imaging
volume experience the same AIF.
27
This also obviates the dilemma of arterial selection (since
different arterial input functions may be obtained in different
representative arteries, leading to different quantitation of
perfusion parameters).
Arterial spin labeling
Another technique offers the promise of quantitative cerebral
perfusion imaging without the need for exogenous contrast agents
(Figure 5). Although not yet in routine clinical practice, the
technique (or rather family of techniques) known as arterial spin
labeling (ASL) merits some discussion, especially in light of
increasing field-strength MRI. It can be applied in situations
where several separate perfusion measurements may be required (eg,
during therapy). Arterial spin labeling techniques are plagued by
nearly as many acronyms as gradient-recalled echo imaging (Table
1), but the techniques can be grouped into two distinct families:
pulsed ASL (PASL) and continuous ASL (CASL).
A PASL sequence is obtained by sending a 180° radiofrequency
(RF) pulse during a very short period of time to a large tagging
slab (approximately 10 cm) just below the circle of Willis,
proximal to the volume of interest, in order to label the water
protons of the arterial blood by inverting their magnetization.
While the blood is flowing into the arterioles and then the
capillary bed, the magnetization of the arterial blood undergoes
longitudinal relaxation. After a certain delay of time when the
blood has transited to the capillary bed and has perfused into the
brain tissue, an image is acquired using a fast sequence, such as
echoplanar or spiral imaging. The experiment is then repeated
without inverting the arterial water protons and the two images are
subtracted to give a map that should show the amount of blood that
has perfused the mass of tissue contained in a voxel during the
time between the tagging pulse and the image acquisition. Since the
absolute signal intensity difference between the two scans
(with/without arterial inversion) is very small, multiple images
are averaged together and the perfusion map is calculated using
models that relate the difference of magnetization to the regional
cerebral blood flow.
28,29
A CASL sequence uses basically the same principle as a PASL
sequence, except that in a CASL sequence the blood is continuously
tagged (for approximately 3 sec) on a much smaller region
(approximately 1 cm) and farther away from the imaging slices (a
gap of approximately 10 cm between the two regions).
Problems and artifacts: Why quantitive ASL is not yet
clinically routine
In fact, ASL entails many problems and artifacts that have to be
reduced in order to realize the promise of accurate flow
quantitation. Tagging of a separate label slice or slab can be
considered as off-resonance excitation for the imaging slice.
Consequently, it may incur magnetization transfer signal loss in
the presence of appropriate macromolecular entities in tissue.
28
Since ASL techniques require a subtraction of a tagged image from a
control one, care must be taken to have the same magnetization
transfer influence during both sequences (ie, the macromolecules
irradiated by the 180º pulse during the tagging sequence must be
similarly affected during the control sequence). This effect is
more important for a CASL sequence because of the long duration of
the RF pulse, whereas it is lesser for the PASL family. Typically
it is addressed by applying downstream or irrelevant 180° RF pulses
analogous to the tagging pulses, during the control image sequence.
30,31
The uncertain transit time for the blood to flow from the
tagging region to the imaging slice makes the quantification
difficult. This may be ameliorated by lengthening the interval
(inversion time [TI]) between tag and image in order to allow the
assumption that all the tagged blood has perfused the slice before
the acquisition. Lengthening TI also allows the arterial tagged
blood to pass through and leave the imaging slice before imaging.
Indeed, tagged blood flowing in arteries passing through the
imaging slice at the time of acquisition would otherwise cause
undesired focal high-intensity regions on the perfusion map. So, a
long TI overcomes two artifacts; however, it also reduces the
intensity of a signal that is already weak (since the tags fade
during T1 relaxation in the limit of TI ~ 5 * T1, tagged blood
appears of course to have the same magnetization as the untagged
blood of the control image, and subtraction of the two images yield
nothing more than noise). It seems that this problem will be
reduced in severity with the implementation of higher
field-strength MR systems (where T1 relaxation times are longer).
32
One also has to consider that the inversion profile of the tag
may not be a perfect box, and, thus, incomplete inversion remains
at the edges of the tagging region. In this way, a sequence called
QUIPSS II (quantitative imaging of perfusion using a single
subtraction) has been developed in order to control the time width
of the tag or, in other words, the amount of tagged blood that
enters the imaging slice.
33,34
To do so, a saturation pulse is applied to the labeling region at
time TI
1
after the application of the label. In this case, there is a
precisely known amount of labeled blood (equal to flow * TI
1
) that leaves the labeled region.
Comparison of DSC and ASL imaging
The most obvious difference between these two perfusion
techniques (Table 2) is the fact that the gadolinium remains
intravascular when the BBB is intact, while the magnetically
labeled water molecules in the blood plasma diffuse relatively
freely into the tissue. With radioactive labeling as used in H
2
15
O-positron emission tomography (PET) and
133
Xe-single-photon emission computed tomography (SPECT) perfusion
scanning, the slow washout of tracer from the tissue back into the
vascular system and the slow radioactive decay of the tag together
define a measurement time-scale that is much slower than the first
pass dynamics of the DSC scan.
35,36
However, in the case of magnetic labeling, the tag decays extremely
rapidly (down to 37% of initial value in TI ~ 1 sec), so that the
situation is dramatically reversed. Whereas the signal loss as the
gadolinium bolus makes its first pass through the tissue's
capillary bed is quite easily monitored using an echo-planar
imaging (EPI) scan, the T1 decay of the magnetic label basically
forces the ASL scan to be performed at a single inversion time (TI
~ 1.4 sec). This time corresponds to the optimal measurement time,
given the competing effects of signal gain over time as labeled
water enters the tissue versus signal loss due to T1
relaxation.
For absolute quantitation of CBF, the model of the tracer
kinetics that is appropriate to the DSC method (nondiffusible
tracer) really requires a local estimate of the AIF that is
relevant to the tissue of interest. Since ASL uses a freely
diffusible tracer, it should not have a fundamental problem with
AIF variability. The equivalent PET and SPECT methods rely on an
integration of the total tracer that is delivered to the different
tissues to form a perfusion map that is independent of the local
AIFs. It is only because the ASL measurement usually has to be
performed before the trailing edge of the bolus of labeled spins
has reached all the tissues of interest that a nominally similar
problem arises. In some of the ASL variants, the trailing edge is
not well-defined, but in the QUIPSS method, a saturation pulse is
applied just below the imaging volume about 0.7 sec after the
inversion pulse.
33,34
This allows another 0.7 sec for the trailing edge of the now
well-defined bolus (with time width = 0.7 sec) to reach all the
tissues within the imaging volume.
From the point of view of systematic errors, ASL has the greater
potential for absolute quantification of perfusion, but the level
of the perfusion signal (approximately 2% of the raw signal) is
extremely low relative to the noise, which will propagate a
relatively large random error into the final CBF map. This is
somewhat improved at higher field strength due to longer relaxation
times, which enable a greater buildup of labeled protons to occur
via the use of a longer inversion time. If the calculation of
ratios of CBV and of MTT between ischemic and relatively normal
tissue is adequate for assessment using DSC, then the main criteria
for choosing between the two methods are the relative speed and
excellent signal-to-noise of the DSC scan. Furthermore, typical ASL
implementations are limited in the number of slice locations
accessible compared with DSC methods, which typically provide 20 to
30 slices or whole-brain coverage. On the other hand, since ASL is
noninvasive and the endogenous tracer literally disappears within a
few seconds, these scans can be used repeatedly to monitor
blood-flow changes during a vascular reactivity test using a
vasodilator, such as acetazolamide or CO
2
.
Reactivity
Ischemic tissue can maintain a reasonable CBF by dilatation of
the precapillary resistance vessels, which causes an increase in
CBV (hence the commonly observed normal or even elevated CBV
measurement, despite ischemia). Eventually, this autoregulatory
mechanism reaches some limit (referred to as the reserve capacity),
and infarction is likely to occur. For a given vasodilatory
challenge, there will be a significant increase in CBF in healthy
tissue but much lower reactivity in tissue with an exhausted
reserve. These studies were pioneered using transcranial Doppler
ultrasound, and have become a fairly standard component of SPECT
perfusion scanning.
37-39
The feasibility of MRI reactivity scans has been demonstrated using
both ASL and blood oxygen level dependent (BOLD) techniques
40,41,42
(Figure 6). The latter uses a T2*-weighted EPI sequence (identical
to that used in DSC imaging) and is sensitive to the susceptibility
effect of deoxyhemoglobin, which is effectively an endogenous
equivalent of the gadolinium used in DSC imaging. Interestingly,
the BOLD susceptibility method also has problems with absolute
quantitation of CBF, since the signal depends on the resting
deoxyhemoglobin level, which depends on resting CBV, CBF, and
oxygen extraction fraction (OEF). Recent studies have shown a
strong correlation between reduced reactivity and increased MTT as
measured by DSC imaging,
43
and both measures are considered to be important in detecting
tissue at risk of infarction.
Conclusion
Both diffusion and perfusion imaging techniques stand at the
forefront of physiologically sensitive radiologic imaging, offering
another tantalizing combination of integrated imaging of anatomy
and function. Although the principal clinical indication is in the
imaging of ischemia, applications are becoming more widespread and
include a broad range of cerebrovascular diseases. Efforts to
extract the maximal clinically relevant information from dynamic
perfusion imaging center both on methods for accurate and precise
quantitation of CBF as well as on intrinsic understanding of the
co-varying behavior of separate perfusion parameters--CBV, delay of
bolus arrival, etc. Surrogate estimates of the degree of collateral
blood supply and distinction between good compensatory
autoregulation and bad loss of vascular tone remain subjects of
active research. The concept of monitoring vascular reactivity (ie,
vascular function, not just vascular state) and indeed vascular
integrity (via the extravasation of initially intravascular
contrast agent, leading to estimates of microvascular permeability)
44
offer considerable promise in the expanding clinical role of these
methodologies.
AR