This publication was supported by an educational grant from
Amersham Health, Princeton, NJ. The opinions expressed in this
publication are those of the authors and not necessarily those of
Amersham Health.
Dr. Bae reports relationships with Tyco Healthcare and
Mallinckrodt through patent agreements and as a consultant. Dr.
Fishman reports relationships with Siemens Medical Solutions and
Amersham Health as a consultant. Dr. Foley reports a relationship
with GE Medical Systems through an investigator agreement. Dr.
Naidich reports a relationship with Siemens Medical Solutions
through its Advisory Board and as a consultant. Dr. Saini reports a
relationship with GE Medical Systems through research support. Dr.
Becker, Dr. Sahani, Dr. Siegel, Dr. Tahktani, and Dr. Zinreich
report that no such relationships exist.
Dr. Foley
directs the Section of Digital Imaging at the Medical College of
Wisconsin, Milwaukee, WI.
As multidetector CT technology has evolved, so has our approach
to contrast administration. This article will discuss the
relationship between technology and contrast delivery, and explore
its implications for pre-intervention planning.
The evolution in CT technology from 4- to 8- to 16-detector
channels has led to an increase in beam width (Figure 1). When
going from a 4-channel scanner (GE Light Speed Plus, GE Medical
Systems, Waukesha, WI) to an 8-channel scanner (GE Light Speed
Ultra), increasing the beam width from 5 to 10 mm doubles the speed
of scanning, assuming detector collimation, rotation speed, and
pitch remain constant. When going from an 8-channel (GE Light Speed
Ultra) to a 16-channel scanner (GE Electric Light Speed 16),
retention of the same 10-mm beam width, coupled with a change in
detector configuration from 8 * 1.25 mm to 16 * 0.625 mm, results
in a doubling of z-axis resolution for the same scan speed.
(Compared with a 4-channel scanner, the same 16-channel
configuration doubles both z-axis resolution and scan speed.)
Alternatively, maintaining detector collimation constant at 1.25 mm
increases the beam width of the 16-channel scanner to 20 mm. In
this case, scan speed is twice as fast with a16-channel scanner as
with an 8-channel scanner, and 4 times as fast as with a 4-channel
scanner.
The general principles governing our approach to multidetector
CT angiography (CTA) are as follows:
* We aim for an arterial enhancement of 250 to 300 HU. One
important consideration is the need to distinguish arterial
enhancement from calcification in the vessel wall, which is
achieved by increasing the window width to 750 and the window level
to 150 on the axial images, with appropriate presets on the
maximum-intensity projection displays.
* In determining cephalocaudad coverage, we take into account
both the need to acquire images during the first
circulation of contrast and the patient's ability to sustain a
breath-hold. We generally consider 20 seconds a reasonable
breath-hold interval.
* To time image acquisition, we inject a mini-bolus of contrast
and observe aortic arrival time. We do not use bolus-tracking
software, as our technology does not allow a sufficiently rapid
transition from the detection of threshold contrast enhancement to
the initiation of scanning.
Aortoiliac CTA
Aortoiliac CTA covers the territory from just above the celiac
axis to the proximal thigh, approximately 30 cm. Figure 2
illustrates the timing of contrast administration and image
acquisition using a 4-channel scanner. With a detector collimation
of 1.25 mm, a pitch of 1.5, and a table speed of 15 mm/sec, the
image acquisition interval is 20 seconds. Contrast material is
injected at
5 mL/sec for 20 seconds, for a total of 100 mL (orange bar). In
this schematic, image acquisition begins at 15 seconds (yellow
bar), as determined by a preliminary mini-bolus, and the
acquisition interval is equal in length to the injection interval.
The injection-to-scan delay will vary from 12 to 30 seconds,
depending on the individual patient's circulation time.
With an 8-channel scanner, z-axis resolution and scan rotation
speed remain the same as with a 4-channel scanner (Figure 3). Beam
width is doubled, however. If at the same time the pitch is reduced
to 1.35, the acquisition interval is reduced to 12 seconds for the
same cephalocaudad coverage. Contrast material is injected at 5
mL/sec for 12 seconds, for a total of 60 mL. Once again, the
linkage between injection and acquisition is determined by the
mini-bolus technique.
Figure 4 shows curved planar reformations of the same patient
studied with a 4-channel and an 8-channel CT scanner on separate
occasions using the same z-axis resolution. Attenuation of the
aorta and iliac system throughout the cephalocaudad coverage is
equivalent, despite contrast volume having been reduced by nearly
half.
With a 16-channel scanner, the best approach to aortoiliac
studies is to maintain the same beam width as with an 8-channel
scanner and decrease detector collimation to 0.625 mm (Figure 5).
If the pitch remains 1.35, the acquisition interval remains 12
seconds, and cephalocaudad coverage is unchanged. The z-axis
resolution is doubled, however. Contrast is injected at 5 mL/sec
for a total of 60 mL.
Figure 6 shows a patient with an aortoiliac stent graft. Images
were acquired on an 8-channel scanner using a slice width of 1.25
mm, and a 16-channel scanner using a slice width of 0.625 mm. The
effect of improved z-axis resolution is clearly evident in the
superior edge definition of the right renal artery.
Stent-graft planning
Important issues in planning for stent-graft surgery include the
ease of iliac access, the transverse dimensions of the superior
neck of the aneurysm, the length of the superior neck (the proximal
stent-graft placement zone), the angle of the neck in relation to
the aneurysm, and the diameter and length of the distal landing
zone in the common iliac arteries.
CT angiography can noninvasively obtain both the length
measurements, which would otherwise be obtained by digital
subtraction angiography with a calibrated catheter, and the
diameter measurements, which would otherwise be obtained by
intravascular ultrasound. Using center-line tracking, seed points,
and automated edge detection, CT models the aorta and iliac
arteries in three dimensions. It provides estimates of the true
vessel length and the diameter, which is measured perpendicular to
the vessel rather than to the body.
CT also plays a key role in postsurgical follow-up and the
evaluation of stent-graft complications. Potential complications
include endoleak (Figure 7), which might result in aneurysm
expansion; graft thrombosis; stent graft migration or kinking; and
branch vessel occlusion.
Renal studies
CT angiography is widely accepted for the evaluation of
potential renal donors. With a 0.625-mm detector collimation, the
16-channel scanner provides detailed information on accessory
vessels, points of branching, and renal artery stenosis (Figure 8).
With a second-pass image acquisition, it can also detect renal vein
anomalies.
In patients with marginal renal function, the 16-channel can be
used as a super 8-scanner by increasing detector collimation to
1.25 mm and thereby doubling scan speed. Under such circumstances,
the acquisition interval is reduced to 6 seconds, and contrast
material is injected at 5 to 6 mL/sec for 6 seconds, for a total
contrast volume of just 30 to 36 mL. One might wonder whether it is
possible to accurately time image acquisition to coincide with the
passage of the contrast bolus, but with the aid of the preliminary
mini-bolus, we have been successful in doing so.
Figure 9 shows an extension of an aortoiliac study into the
upper thigh of a patient with bilateral renal artery implants,
bilateral internal ureteral stents, a femoral-femoral artery
crossover graft, and serum creatinine level of 1.5 to 2 mg/dL. This
study was done with 40 mL of contrast material.
Table 1 summarizes the comparison of 4-, 8-, and 16-channel
scanners for aortoiliac CTA.
Thoracoabdominal aortic CTA
There are two approaches to imaging the thoracoabdominal aorta
with the 16-channel scanner. Detector collimation can be set at
1.25 mm for a full 20-mm beam width. In this case, the acquisition
interval for the thoracoabdominal aorta is equivalent to that for
an aortoiliac study, despite a cephalocaudad coverage of 60 cm.
Contrast volume is approximately 60 mL. The alternative approach is
to use a detector collimation of 0.625 mm. In this case, both the
acquisition interval and the contrast load will double.
The choice between the two ap-proaches is guided by the
patient's cardiovascular status and renal function, as well as the
indication for the study. For pre-operative planning in a patient
with a suspected thoracoabdominal aneurysm, it is essential to
obtain detailed information on small vessels that parallel the
slice plane. Therefore, it would be appropriate to select a
detector collimation of 0.625 mm and a contrast volume of 120 mL.
In the case of aortic dissection, where small-vessel detail is less
important, a detector collimation of 1.25 mm is sufficient, and
contrast volume can be reduced to 60 mL.
Extremities
Patients are usually referred for CTA of the extremities in
preparation for vascular or plastic surgery, or for evaluation of
orthopedic problems. Multidetector CT is both suited for and
challenged by the demands of imaging the extremities.
The distance from the renal vascular pedicle to the feet is
approximately 120 cm. To cover this distance with a 4-channel
scanner, it is necessary to use a 2.5-mm detector collimation, a
pitch of 1.5, and a table speed of 3 cm/sec, resulting in a
40-second acquisition interval. Contrast material is injected at 4
mL/sec for 40 seconds, and a mini-bolus of contrast material
determines the timing of image acquisition.
Lower-extremity angiography becomes more complex in patients
with very slow runoff, particularly below the knee. Under such
circumstances, we delay initiation of image acquisition beyond the
aortic arrival time as determined from the contrast mini-bolus, in
order to avoid scanning faster than the contrast circulation.
With an 8-channel scanner, we reduce detector collimation to
1.25 mm. With no changes in the other scan settings, the image
acquisition interval remains about the same, 44 seconds. With a
16-channel scanner, we generally maintain detector collimation at
1.25 mm, which reduces the acquisition interval from 44 to 22
seconds. We inject contrast material at 5 mL/sec, for a total
contrast volume of 110 to 120 mL. This volume of contrast material
is acceptable in patients with marginal renal function or
substantial cardiovascular disease.
Table 2 summarizes differences in the scan protocols for CTA of
the extremities using 4-, 8-, and 16-channel scanners.
A key issue in lower-extremity CTA is the need to determine the
degree of stenosis in areas with significant calcified plaque. This
problem necessitates the use of curved planar reformations and
automated measurement techniques, to determine the diameter and
area at the sites of stenosis (Figure 10).
Image processing remains a key issue, particularly in extremity
CTA. It is important, for example, to be able to do bone
segmentation at the workstation effectively and quickly. Vessel
tracking, particularly below the knee, and automated stenosis
sizing are also critical in making CTA of the extremities a robust
and acceptable technique that will replace other forms of
angiography.
Conclusion
Intravenous CTA of the thoracoabdominal aorta and abdominal
visceral vessels is a robust technique that provides all the
requisite diagnostic information for pretherapy planning and
postintervention follow-up. Improved performances associated with
the progression from 4- to 8- to 16-channel scanners is, in each
instance, associated with either faster acquisition, improved
resolution, or reduced contrast load. Intravenous CTA of the
extremities can be used for preoperative evaluation of iliac,
femoral, and popliteal arterial disease. Display of isolated tibio
peroneal arterial disease can be more problematic.
Figure Captions
FIGURE 1.
Matrix detector technology evolution. The beam width of an
8-channel CT scanner, at 10 mm, is twice as large as that of a
4-channel CT scanner. Detector collimation, at 1.25 mm, remains
unchanged, resulting in a doubling of scan speed. With a
16-channel scanner, the beam width can remain 10 mm while
detector collimation can be reduced to 0.625 mm, doubling z-axis
resolution. Alternatively, beam width can be increased to 20 mm
while detector collimation remains 1.25 mm, doubling scan speed
in comparison to an 8-channel scanner, or quadrupling it in
comparison to a 4-channel scanner.
FIGURE 2.
Timing of contrast administration and image acquisition during
aortoiliac CTA using a 4-channel scanner. Detector collimation is
1.25 mm, pitch (Z) is 1.5, and table speed 15 mm/sec (7.5 mm per
scan rotation/0.5 sec per rotation). The image acquisition interval
is 20 seconds. Contrast material is injected at 5 mL/sec for 20
sec, for a total of 100 mL (orange bar). Image acquisition begins
15 seconds after the beginning of contrast injection (yellow bar),
as determined by a preliminary mini-bolus for this hypothetical
patient. (Note: The injection-to-scan delay will vary from 12 to 30
seconds, depending on individual patient
'
s circulation time.) The acquisition interval is equal to the
injection interval.
FIGURE 3.
Timing of contrast administration and image acquisition during
aortoiliac CTA using an 8-channel scanner. Scan rotation speed
and z-axis resolution are the same as with a 4-channel scanner.
Beam width is doubled, however. If pitch (Z) is reduced to 1.35,
the acquisition interval is reduced to 12 seconds. Contrast
material is injected at 5 mL/sec for 12 sec, for a total contrast
volume of 60 mL.
FIGURE 4
. Curved planar reformations from the same patient studied
with both (A) a 4- and (B) an 8-channel CT scanner on separate
occasions using the same z-axis resolution. Attenuation of the
aorta and iliac system throughout the cephalocaudad coverage is
equivalent, despite contrast volume having been reduced by nearly
half.
FIGURE 5.
Timing of contrast administration and image acquisition during
aortoiliac CTA using a 16-channel scanner. Reducing detector
collimation to 0.625 mm doubles z-axis resolution. Acquisition
interval is 12 sec, and contrast material is injected at 5 mL/sec
for a total of 60 mL.
FIGURE 6.
Patient with an aortoiliac stent graft. Images were acquired on
(A) an 8-channel scanner using a slice width of 1.25 mm, and (B)
a 16-channel scanner using a slice width of 0.625 mm. The effect
of improved z-axis resolution is clearly evident in the superior
edge definition of the right renal artery in the image acquired
on the 16-channel scanner.
FIGURE 7
. (A) Curved planar reformation of a patient with a Type 2
endoleak (arrow). (B) Due to graft limb thrombosis (large arrow),
the patient has had a femoral-femoral bypass graft (small
arrows).
FIGURE 8.
Potential renal donor. With a 0.625-mm detector collimation, the
16-channel scanner can provide detailed information on accessory
vessels, points of branching, and renal artery stenosis.
FIGURE 9.
Extension of an aortoiliac study into the upper thigh of a
patient with bilateral renal artery implants, bilateral internal
ureteral stents, an axillofemoral and femoral-femoral artery
crossover graft, and serum creatinine level of 1.5 to 2 mg/dL.
This study was done with 40 mL of contrast material.
FIGURE 10.
(A) Maximum-intensity projection shows heavy calcification in the
distal aorta, iliac arteries, and the femoral arteries down to
the knee. (B and C) Curved planar reformations show the internal
lumens of the right and left iliac, femoral, and popliteal
arteries without obscuration by the heavy arterial wall
calcification. The left superficial femoral artery is occluded
and distal runoff is via the left profunda femoris artery with
collateral reconstitution of the left popliteal artery.
Discussion
ELLIOT K. FISHMAN, MD:
Thank you, Dennis. Any questions?
KYONGTAE T. BAE, MD, PhD:
You said you like to have desired arterial enhancement of 250 to
300 HU. But then with some of the reduced contrast volume, you are
not going to achieve that, right?
FOLEY:
That's a good question. The issue relates to the fact that we are
not really dealing with recirculation and humping up the aortic
profile. So if you have a very short acquisition of 6 seconds, you
don't get the recirculation effect. I haven't gone systematically
to look at those patients, but they probably don't have as much
aortic attenuation as we have been talking about.
BAE:
Just from looking at the image, it looks like enhancement is not at
the level of 250 HU.
FOLEY:
Yes, but the compensation for that is--instead of injecting at 5
mL/sec--inject at 6, and maybe inject at 7 mL/sec. One thing I
didn't indicate on that schematic is that, actually, when we go to
a 6-second acquisition we increase the in-jection rate to 6 mL/sec
instead of 5.
SANJAY SAINI, MD:
Also, when you have a longer coverage, why not go to a 1.75 pitch?
You stayed at 1.35.
FOLEY:
That's a good point. With regard to the vessel detail, if you're
looking at a vessel that's perpendicular to the slice plane, it
probably would not actually diminish anything at all, that would be
an advantage.
FISHMAN:
In terms of CTA, outside of the heart, the most challenging is
definitely the lower extremity in terms of timing. You said that if
you were using the automated bolus tracking, you measured the lower
abdominal aorta. What values do you use for that?
FOLEY:
Actually, we are still using the mini-bolus technique. So we don't
use the thresholding. I'd be interested in other people's
opinions.
FISHMAN:
But even with the mini-bolus, there is the issue. Trauma patients
with glove/ring injuries are easy to image. When you evaluate a
60-year-old for atherosclerotic disease, what's good for the left
is not good for the right. What do you do in those situations?
FOLEY:
We haven't had a sufficient experience to determine that.When you
have iliac occlusive disease or femoral occlusive disease,
generally the collateral circulation that you fill out with an
intravenous injection overcomes the problem that you have if you do
selective arterial injections with digital subtraction angiography.
But you're right, below the thigh and below the knee, it's going to
be an issue. You just hope that you have such a long bolus, and
you've got some latitude between a long bolus and a long
acquisition interval, that you're not going to run into a
problem.
I've heard other people present material on CTA of the lower
extremities who have not really indicated there's a major problem
with different flow rates between the two sides.
FISHMAN:
But, you did a lot of angiography, and you know from that there's a
significant problem. We do sort of a crude thing: I just watch the
images in all those CTAs. If I see that the left looks good and the
right doesn't, I just wait 20 more seconds, and scan again. I think
it's impossible for patients to be perfect on both.
CHRISTOPH R. BECKER, MD:
What would be a good idea, I've heard, is to test bolus in the
femoral artery, to see when the arrival is there, and to then
concentrate on the diseased leg.
FOLEY:
Yes, that's a good point.