Dr. Schwarz
is an Associate Professor of Medicine and the Director of the
Echocardiography Laboratory at University of Rochester Medical
Center, Rochester, NY.
Microbubble echo contrast is well-documented to improve left
ventricular endocardial border delineation in patients with
technically limited examinations, especially when harmonic imaging
is used. However, it is not uncommon for some patients with very
technically limited exams to have less than complete endocardial
visualization, even with the use of harmonics and echo contrast.
Such may often be the case for stress echo patients in the
immediate post-stress period. Harmonic tissue Doppler energy (TDE)
combines the motion filter of tissue Doppler (which rejects
non-moving image artifacts) with the image improving effects of
harmonic imaging (higher resolution imaging and near elimination of
side lobe artifact compared to fundamental frequency imaging) and
will often improve image resolution and dramatically reduce image
clutter. In addition, high-power tissue Doppler ultrasound enhances
microbubble imaging through bubble destruction, which creates
broadband ultrasound transients and results in loss of correlation
(for the Doppler autocorrelator) and a stronger than normal Doppler
signal. Combining harmonic TDE as a semi-transparent overlay to the
two-dimensional (2D) native tissue harmonics image (NTHI) maximizes
the contrast enhancement of the blood-pool relative to the
myocardial tissue. In this way, the effects of Harmonic TDE and the
NTHI are additive to improve endocardial border delineation, while
at the same time not increasing image clutter. Image frame rates of
at least 30 Hz are possible using ultrasound scanners containing a
coherent beamformer, thus allowing real-time Harmonic TDE imaging
of endocardial motion with similar temporal resolution to
traditional 2D NTHI.
Background
The sources of image clutter, and how NTHI and TDE imaging
improve ultrasound images in technically limited exams with a low
signal-to-noise ratio, will be discussed in detail in a forthcoming
clinical white paper.
Echo Contrast Mini Bubble Physics Review
All echo contrast agents currently available consist of very
small microbubbles delivered intravenously in an aqueous medium.
These microbubbles initially exist entirely in the blood-pool and
enhance ultrasound backscatter through two primary mechanisms: 1)
speed of sound change at the gas-fluid interface of the
microbubble; and 2) though micro-bubble oscillation, which
increases the effective cross-section of the micro-bubble and also
radiates ultrasound energy. The optimum steady-state backscatter is
achieved from a microbubble when it is insonified at its resonance
frequency with optimal power settings. The resonance frequency is
based on a number of factors, but the most important are the bubble
size and the insonification frequency. The resonance frequency can
be roughly calculated as the dividend of 6 and the bubble diameter.
At standard clinical harmonic frequencies (transmit 2 to 1.75 MHz),
this corresponds to bubble diameters of 3 to 3.5 µm. The echo
contrast agents currently available for clinical use consist of a
population of bubble sizes, but the population means of bubble
diameters are in this range. A higher transient spike of ultrasound
backscatter may also be elicited from bubbles when they are exposed
to higher power ultrasound and there is microbubble destruction.
Thus, optimal steady-state contrast imaging may be achieved at the
resonance frequency using modest insonification energies, but
maximal contrast imaging will occur using high-power insonification
energies at the resonance frequency. High-power insonification
energies destroy the bubbles as they are being imaged, so rapid
replenishment of contrast is required to achieve continuous
imaging.
How Echo Contrast Improves Left Ventricular (LV)
Endocardial Border Delineation
Imaging of the LV endocardial borders are limited by factors
that attenuate the ultrasound beam (obesity, narrow intercostals
spaces, hyperinflated lungs in pulmonary disease patients, etc.)
and cause cardiac displacement (as in obese patients and those with
pulmonary disease). LV endocardial visualization is also limited by
the lateral resolution of the ultrasound system, because much of
the endocardial border is parallel to the ultrasound beam when
imaging from the apical window. Microbubbles are the solution to
these problems based in part on their shape (spherical microbubbles
are always imaged with axial resolution) and by their enhanced
backscatter (especially at the harmonic frequencies) compared with
the non-bubble surrounding medium. When given in sufficient
quantities, the borders of the microbubble-rich blood-pool define
the endocardial border.
How 2D Harmonic Imaging Improves Microbubble Imaging
Second harmonic imaging improves the imaging of microbubble echo
contrast in part the same way it improves NTHI. Compared with
fundamental frequency 2D imaging, harmonic imaging uses relatively
high receive frequencies and a more narrow transmit beam (better
image resolution) and NTHI transmit beams have less side-lobe
content (less image clutter). Microbubble backscatter power is
greatest at the fundamental frequency. However, the relative
backscatter power of microbubbles compared with tissue is much
greater at the second harmonic frequency than at the insonification
frequency (Figure 1). This leads to improved border delineation and
better detection of low concentration microbubbles, when using
second harmonic imaging (Figure 2). In addition to these factors,
high-intensity imaging produces violet microbubble oscillations and
sudden bubble destruction, which produces transients of even
greater harmonic backscatter.
How TDE Reduces Improves Microbubble Imaging
TDE imaging improves NTHI imaging for the detection of
endocardial borders through the addition of the Doppler motion
filter (removes unwanted non-moving image "clutter"). As noted
above, microbubbles that are oscillating violently and are
destroyed in high intensity sound fields transmit high-intensity
broadband transient ultrasound bursts. The Doppler auto-correlation
algorithm interprets the ultrasound transients associated with
bubble destruction as motion, usually of uncertain velocity, but of
high intensity. This effect typically creates a vivid color Doppler
image with marked "aliasing" noted, even if there is no blood flow
velocity. The tissue Doppler energy map also produces a vivid
image, but it appears uniform because the energy map is devoid of
velocity information. The endocardial border is identified as the
border zone between the bubble-rich blood-pool and the bubble-poor
myocardium. TDE has much higher frame rates than traditional color
Doppler due to a smaller packet size, especially when a coherent
beamformer is used. This allows TDE imaging of all but the largest
ventricles at a frame of 30 Hz or more.
Combined NTHI and TDE for Microbubble Imaging
TDE images are displayed as a semitransparent overlay to the
NTHI image. Both TDE and NTHI enhance the bubble-rich blood-pool
relative to the bubble-poor myocardium. Therefore, bubble-rich
backscatter is enhanced when NTHI and TDE are displayed as an
additive image, when compared with displaying either singularly
(Figure 3).
Instrument Settings for NTHI/TDE
The goals of the instrument settings for NTHI and TDE are to
maximize the contrast-rich intracavitary blood-pool backscatter,
while at the same time minimizing the bubble-poor myocardial
backscatter. Stated another way, the left ventricular cavity should
appear "white" while the myocardium should appear "black." This is
exactly the opposite of the normal goals when imaging the left
ventricle without contrast; normally the machine is adjusted so
that the myocardium is white and the cavity is black. As noted
above, both contrast and tissue backscatter intensity is similar
when imaged at the insonification frequency (Figure 3). This leads
to a "white" left ventricular cavity and a "white" myocardium (poor
differentiation between contrast and tissue). However, tissue
backscatter is much lower in intensity relative to contrast bubbles
at the second harmonic, and this difference can be accentuated if
low transmit power settings are used. This is due to the nonlinear
relationship between transmit power and NTHI backscatter; for every
drop in transmit power, there is a correspondingly larger drop in
NTHI backscatter. In contrast, microbubble resonance still occurs
at lower power, which accounts for the much greater backscatter
intensity of microbubbles relative to tissue at low transmit powers
when using harmonic receivers.
The instrument settings for NTHI and TDE are best adjusted
separately, usually before the introduction of echo contrast. Each
patient will require minor adjustments to the settings as described
above, depending on the imaging circumstances. Occasionally, the
settings will need to be adjusted during the contrast infusion. In
the case of stress echocardiography, the pre-stress image settings
usually determine the post-stress settings and it is seldom
necessary to make adjustments to the instrument in the post-stress
period.
Contrast 2D Harmonic Settings (Table 1, Figures 4A and
4C)
Normally maximal transmit power is used for noncontrast NTHI to
maximize the amount of nonlinear wave propagation and thus maximize
NTHI backscatter. The 2D harmonic settings are slightly different
when contrast is employed compared with harmonic imaging used
without contrast. The main difference is that the output power is
adjusted downward to maximize bubble backscatter relative to tissue
and to minimize the bubble destruction associated with high output
power. Typically, transmit power settings of ¾0.8 MI are used for
harmonic contrast imaging. As is the case in NTHI imaging, the
transmit focus should be placed at the area of interest for small
targets or near the far-field boundary for larger targets (ie, near
the mitral valve plane, when imaging the LV from the apical
window). The image dynamic range needs to be set at a level that
places the contrast bubble echoes in the middle of the image
gray-scale, while at the same time minimizes the display of
myocardial echoes. This is usually achieved with a dynamic range
setting of about 60 dB, with the image (log) compression curve
chosen to further suppress low intensity (myocardial) backscatter
and image noise. The optimum gain setting is one in which
endocardial borders are just barely visible on the pre-contrast
images and without excess clutter artifact. This may mean that the
NTHI image looks quite light (under-gained). Regardless of the
actual overall gain setting, it is critical
that the depth-dependent time-gain (TGC) slider controls be used
for adjusting gain first, with overall system gain being used only
for final adjustments. In all cases, the TGC slider controls should
be somewhere in the 50% to 75% gain positions with the overall
system gain used to bring the NTHI image either up or down in gain
to the optimal range.
Contrast Harmonic TDE Settings (Table 1, Figures 4B and
4D)
The instrument is set to color TDE imaging and adjusted so that
the imaging frequency is the same as that used for NTHI imaging
(typically transmit 1.75 MHz, receive 3.5 MHz). The frequency
adjustment is important so that the instrument can be optimized in
both the NTHI and TDE modes for simultaneous use. The transmit TDE
power of the machine is set to maximum, so that there will be
maximal microbubble oscillation, including some microbubble
destruction, which will enhance the color Doppler signal. The color
filter should be set to give good sensitivity while removing
stationary "clutter" from the image (filter 2 on the Acuson
products). The harmonic color TDE dynamic range is typically
adjusted to a level about the same as the NTHI image, but usually
not more than 60 dB and not less than 50 dB in very technically
limited cases. The NTHI image then needs to be turned off
temporarily, leaving just the harmonic color TDE image. Assuming
the machine has already been setup for optimal NTHI imaging (TGC
slider controls in the 50% to 75% gain positions), Doppler gain is
then adjusted to just bring out endocardial borders without
excessive image artifact. This gain setting is less than typical
for noncontrast imaging so that the system will not be over-gained
once contrast is added to the image. The image frame-rate is
optimized through a combination of reducing 2D harmonic/TDE image
width, and by adjusting TDE packet-size and line-density
("Space-Time" control to T2 on the Acuson products). A frame-rate
of at least 30 Hz should be achievable in nearly all cases when
using an instrument with a coherent beamformer. The TDE image is
then changed back to the semi-transparent overlay with the NTHI
image (mix = 3 or 4 on the Acuson products) and a pleasing color
map is chosen. While any color map will suffice, a gray-scale map
is often used so that the combination 2D harmonic/harmonic color
TDE image appears no different from a more traditional 2D NTHI
image.
Combined 2D Harmonic/TDE Settings
Echo contrast should appear as a very bright substance that
fills the entire LV cavity while the myocardium looks quite dark,
thus allowing for the identification of the endocardial boundaries.
No adjustments of the instrument settings should be required during
contrast imaging, if the machine has been setup as described above.
It is not uncommon for "swirling" of the contrast to be observed in
the near field of the ultrasound image, where the ultrasound
intensities are highest. The "swirling" is due to a combination of
microbubble destruction and movement of the bubbles due to acoustic
radiation force, and usually most obvious in the LV apex when
imaging from the apical window. The appearance of "swirling" is
dependent on the ultrasound transmit intensity, the fragility of
the microbubbles, and on the rate of microbubble replenishment in
the imaging area. It is for this reason that low transmit
intensities are used for 2D harmonic contrast imaging, but high
transmit power is required for TDE imaging. Under resting
conditions, the rate of contrast replenishment in the imaging area
is dependent on the rate or dose of contrast administered.
Therefore, it is typical to use a higher rate of contrast infusion
when using TDE imaging compared with 2D harmonic imaging alone. In
the case of peak or post-peak stress imaging, the rate of contrast
replenishment is using high enough to avoid "swirling" due to the
high cardiac output and no adjustment to the contrast infusion rate
need be made.
Potential Uses for Contrast 2D Harmonic/TDE
Imaging
Contrast 2D harmonic/TDE imaging can be used in all settings
that might benefit from contrast enhancement of the blood-pool,
regardless of the pre-contrast image quality. While the addition of
TDE results in an improved border delineation between the
bubble-rich blood-pool and tissue compared to 2D harmonic imaging
in patients with high-quality acoustic windows, the benefits of TDE
are most apparent in patients with low-quality acoustic windows. In
this setting, the addition of the TDE motion filter combined with
the high-intensity acoustic transients associated with high
transmit power induced microbubble destruction, results in greatly
improved signal-to-noise ratio compared to 2D harmonics alone. It
is not uncommon to have diagnostic contrast 2D harmonic/TDE images
in patients with nondiagnostic contrast 2D harmonic imaging used
alone. Therefore, the target patient population most likely to
benefit from the combined imaging approach is patients with very
limited noncontrast 2D harmonic images that need resting or stress
evaluation of LV systolic function. *