Almost one year ago, Dr. Henry Wagner, in this space, presented
very early data on gamma camera coincidence imaging in several
different entities. During the ensuing year, a good deal more
information has become available as over 70 coincidence cameras are
in the field. The purpose of this article is to provide an update
for the practicing radiologist and nuclear medicine physician on
the current capabilities of coincidence imaging on a gamma
camera.
Principles
The principles of coincidence imaging are well known and have
been available in medicine for many years. The original positron
cameras existed as long as 30 years ago.1 Beginning in the 1970s,
significant research time and effort was devoted to the development
of an imaging device that could effectively display the
distribution of positron radiopharmaceuticals in patients. The
result of this effort was the positron emission tomographic scanner
(PET).2 This device takes advantage of the annihilation radiation
emitted when a positron and electron fuse and annihilate one
another. From this interaction, two 511 keV photons are emitted
that are 180° apart. By timing the interaction of photons in
opposing detectors, coincident events can be identified. Though
some non-coincidence events contain useful information, the focus
of coincidence detection has always been to deal as much as
possible with coincidence events.
As it was recognized early on that counting statistics were
crucial in coincidence imaging, special crystals were developed for
dedicated PET scanners. These crystals permitted a much higher
sensitivity than could be obtained with sodium iodide, but they
also decreased the resolution of the system while increasing its
cost. The cost of a dedicated PET scanner, despite its apparent
clinical efficacy, has limited the widespread introduction of these
devices, with only about 60 being used in the United States at
present.
The development of the sodium iodide-based PET scanner at the
University of Pennsylvania in the late 1980s renewed interest in
the use of sodium iodide as a detector system for coincidence
events.3 The cost of this device was lower than the standard PET
scanner, though its sensitivity was lower as well. However, due to
a higher energy resolution than dedicated PET devices (12% for the
camera versus 30% for dedicated PET scanners) it demonstrated
decreased scatter acceptance. In retrospect, perhaps the major
impact of this device was to raise considerations as to whether
standard gamma cameras using sodium iodide detectors could be
acceptable devices for coincidence detection.
Sodium iodide gamma camera as a coincidence
device
Over the last several years, significant attention and research
has been dedicated to optimizing the performance of the standard
gamma camera in the coincidence mode. Several important issues have
emerged with regard to whether or not gamma camera detectors can
adequately serve as coincidence devices. In order to look at this
issue, we must consider some of the problems associated with the
sodium iodide detector.
Over the last 25 years, the gamma camera has progressively been
optimized for low energy imaging. Thinner crystals have been
developed and low-energy collimator systems have been introduced.
Detector energy and linearity mapping have been optimized so that
low energy isotopes are better imaged. The thinner crystals
intended to improve low energy imaging have sacrificed sensitivity
for higher energy isotopes.
A major factor in dealing with coincidence events is the
sensitivity of the gamma camera detector at 511 keV. In order to
improve the sensitivity, the option of thicker detectors has been
employed. Currently, commercially available systems have sodium
iodide detectors ranging up to 3/4" in thickness. However, detector
thickness as great as 1" have been considered and may ultimately
prove desirable in this application. The optimal sodium iodide
crystal thickness in a theoretical situation might be as much as
2.5 inches, but practical considerations may well limit units to
thinner crystals.
A potential implication of using thicker crystals is decreased
resolution at other lower clinical energies. While this would have
been enough to halt further consideration a number of years ago,
the more recent development of resolution recovery algorithms for
use in digital gamma cameras has made the utilization of thicker
crystals a more attractive option.
Another issue, that of detector count rate, is one that is
somewhat misunderstood. There are effectively no collimators in
place in the coincidence mode. Because the collimator typically
screens up to 99.99% of the photons that would impact the crystal,
the count rates for coincidence studies can be remarkably high
without a collimator. When acquiring a patient study at high count
rates, one must be concerned with the dead- time losses and
potential paralysis of the detector. Typically, in an FDG study
begun one hour after administration of
5 mCi of FDG to the patient, the count rate is on the order of
1.3 million per second. If the detector is not capable of handling
this count rate, then artifacts in the clinical images may be
generated.
If one attempts to do a measured attenuation correction by
transmission imaging of the patient, using radioactive sources
mounted on opposing detectors to generate transmission data, the
count rate will increase further. While the impact of measured
attenuation correction will be discussed below, the detector
requirements are further increased with regard to count rate when
this option is employed. Because lower counts in the study create a
major difference when comparing dedicated PET and gamma camera
systems, one should investigate systems that yield the highest
trues (true coincidence events) rates.
Attenuation correction
Before considering attenuation correction for coincidence
imaging, a few observations should be made relative to single
photon imaging. Attenuation correction for single photon studies
has received a great deal of attention in recent years.
Particularly where cardiac single photon emission computed
tomography (SPECT) is involved, many users feel that correcting for
body thickness and intervening structures will decrease or
eliminate inferior wall and breast artifacts.4,5 Clinical
attenuation correction of SPECT studies is common. However,
problems exist with regard to methods that do not employ measured
attenuation. In measured attenuation correction, a transmission
study of the patient is performed and the attenuation correction
maps are generated from that data. Unfortunately, when estimating
correction by computing the thickness of the patient and applying
the linear attenuation coefficient, errors tend to result.
The chest is a particularly complex area for attenuation
correction. Multiple attenuating structures exist. These include
water density, bone density, and air density structures. The best
method that has been found to accurately compute attenuation for
the chest is to measure it by transmission imaging.6 As noted
above, when attenuation correction is applied to coincidence
imaging, detector performance becomes critical.
Is attenuation correction necessary?
On initial thought, it would seem that attenuation correction
for a 511 keV photon would be unnecessary. This high-energy
emission should be less affected by attenuation. However, for
coincidence imaging, attenuation correction is actually more
important than for many single-photon applications. The combined
path length of the two gamma rays generated from an annihilation
interaction is twice that of a single gamma ray of the same energy
arising from the same location. Attenuation is therefore magnified
in this setting.
This debate over attenuation correction has been going on for a
number of years in both the PET community and, more recently, in
the coincidence camera community. There are advocates who present
data on both sides of the question. For PET, the effort required in
the most common methodology makes attenuation correction a
cumbersome procedure. In the gamma camera community, however, the
procedure is highly automated and adds no more than five minutes to
the data acquisition and two minutes to the processing. Routine
attenuation correction of coincidence data on the gamma camera is
now available.
With such a small investment in time and money, there is
increasing interest in the use of attenuation correction on the
gamma camera. In a study of 25 patients in our laboratory, data was
reconstructed from the same data set both with and without measured
attenuation correction. Three experienced nuclear medicine
physicians were asked to review the data and grade it in regard to
diagnostic quality, the presence of artifacts, and overall image
quality. The attenuation-corrected images consistently graded
statistically significantly higher than the non-attenuation
corrected images, particularly in the region of the chest. Lesion
visibility, location of the lesion, and the general accurate
representation of isotopic distribution were improved after
attenuation correction (figure 1). In addition, five lesions were
identified that were not seen on the non-attenuation-corrected data
or were felt to be part of a normal organ on the
non-attenuation-corrected images (figure 2).
Clinical applications
There are a number of clinical applications that have been
advocated for gamma camera coincidence imaging. Predominantly,
these are based on information developed in the PET literature.7
The first of these studies to accrue any significant number of
patients has been the multi-center cooperative trial on imaging
lung cancer. In this protocol, patients who were surgical
candidates or mediastinoscopy candidates were studied with
coincidence imaging on a gamma camera. Some of these patients also
were studied by dedicated PET scanner. The results of the study
were correlated with the surgical/pathologic findings. To date,
approximately 150 patients have been studied in this protocol.
When the correlation was made between independent reading of the
coincidence images and the pathologic data, a sensitivity of
approximately 94% was obtained; specificity in the same patient set
was approximated at 84%.8 About one-half the lesions in the group
were smaller than 3 cm in size. This data is comparable to
previously published data for dedicated PET scanners.
In a small group of patients, success has been reported in
detecting malignant lymphoma with coincidence imaging.7 At the
moment, one of the most significant applications at our institution
is the use of the FDG study for the growing lymphoma population. As
long time advocates of gallium-67 for imaging of suspected
lymphoma, we had the opportunity at our institution to look at the
crossover in a limited number of patients. While gallium remains a
viable lymphoma imaging radiopharmaceutical, FDG appears to avoid a
number of the problems inherent in gallium-67 imaging. For example,
intraluminal bowel activity is not identified on the FDG study.
However, when imaging the pelvis it is necessary that the urinary
bladder be drained during the examination. In one of our pediatric
patients with non-Hodgkin's lymphoma, regrown thymus raised
questions as to the presence of disease on the gallium scan as
residual disease (figure 3). The FDG study in that patient was
normal. At biopsy, only normal thymus was identified.
There appears to be great promise in the use of FDG for the
detection of recurrent lymphoma. Again, in anecdotal situations, a
number of patients have been identified who demonstrate more
extensive disease on the FDG study than on the CT scan. At least
two patients at our institution have been restaged purely on the
basis of FDG. However, a more systematic appraisal of the role of
gamma camera FDG in lymphoma is required based on this early
data.
Other areas where FDG imaging is of interest include breast
imaging and ovarian cancer. At the present time, there is
insufficient data to support the routine use of gamma camera
coincidence imaging in either of these applications. However,
research studies are under way in an effort to further determine
the accuracy in both of these conditions. Ovarian cancer is of
special interest to the oncology community because the detection of
recurrence is extremely difficult by standard means. As of this
writing, however, we cannot routinely advocate the use of FDG
coincidence imaging in either breast cancer or ovarian cancer.
Recurrent brain tumor studies
As a relatively small homogeneous organ, brain imaging on a
coincidence gamma camera is much less fraught with technical issues
than the imaging of chest lesions. Attenuation correction is less
of a problem in the brain, and the major consideration for both
dedicated PET and gamma camera PET becomes the appropriate
correlation with other imaging modalities to precisely localize FDG
uptake. As "fusion imaging" continues to develop, there is every
reason to believe that the overlay of CT and MR data with FDG will
improve the diagnostic accuracy of all these methods in the patient
with recurrent brain tumor.
Cardiac FDG studies
Estimates at our institution have indicated that up to 15% of
patients undergoing cardiac revascularization do so with unclear
indications. In some patients, the decision as to whether to
perform a heart transplant or revascularize the heart is difficult.
In these cases, FDG cardiac imaging may prove of use. For example,
discovery of a hypoperfused section of myocardium that has
significant FDG concentration may push the clinician to try
revascularization for this patient, if it is technically feasible,
rather than transplantation.
Collimator-based FDG cardiac studies appear to offer information
that is equivalent to that of coincidence images. However, they do
not deal with issues of attenuation correction. In collimated
studies, attenuation correction is less of a problem than in the
coincidence mode (figure 4).
Collimator FDG studies
As an interim step in the evolution of FDG imaging using the
gamma camera, high energy collimators have historically been
employed to image FDG. Favorable results have been attained for
cardiac studies, although attenuation correction usually was not
employed. However, some problems are encountered when attempting to
apply collimator FDG to oncology applications. Perhaps the most
significant issue is the low count sensitivity encountered in
collimator studies. If the argument is made that gamma camera
coincidence imaging has decreased count sensitivity, as compared to
dedicated PET, the collimator studies have an order of further
magnitude decrease in count sensitivity. Because the acquisition
time cannot be lengthened in order to try to overcome this issue,
due to the short half-life of FDG, collimated studies remain count
poor in comparison to coincidence detection.9
Reimbursement
In the ideal scenario, reimbursement would not affect the choice
of technology to be employed for imaging. In
the real word, however, this often
has not been the case. The cost of FDG and coincidence devices
was hard to justify when no reimbursement was available. However,
with the recent decision by the Health Care Financing
Administration (HCFA) to reimburse for coincidence imaging related
to staging of non-small cell lung cancer and evaluation of solitary
pulmonary nodules, and a willingness to look at other applications,
a major barrier to the introduction of these devices has been
lowered.
Summary
Judging from the current trends, it is becoming clear that gamma
camera coincidence imaging will play a significant role in nuclear
medicine's future. In order to properly utilize these devices,
however, an understanding of the biodistribution of FDG and the
physics of gamma camera coincidence operation are imperative. The
performance of these devices, and the additional diagnostic
information they bring to bear in various clinical settings is
starting to appear. While debate still exists over the method and
the necessity for attenuation correction, our clinical experience
indicates that attenuation correction improves image quality and
diagnostic yield.
The number of coincidence gamma cameras from all vendors is
likely to exceed the number of PET scanners in this country in a
short period of time, bringing the benefits of FDG imaging to the
community at large. Gamma cameras should not be held to a
comparison with dedicated PET. Rather, they should stand on their
own in determining whether this technology offers sufficient
additional clinical information to be routinely useful. In the
study noted above on non-small cell lung cancer, coincidence gamma
cameras would appear to meet this criterion. When clinical FDG data
is not available, management decisions often are made without full
knowledge of the status of the patient.
Finally, there is the question of timing. Should you wait for
the optimal device or start FDG imaging immediately? Analogous
areas exist elsewhere in medical imaging. For example, computed
tomography (CT) studies are performed on a wide variety of
instruments. While spiral CT may be the current state of the art,
CT examinations are not denied to patients when there is no spiral
CT locally available. Non-spiral examinations still provide
information that can be used in patient management. Dedicated PET
scanners may not be available in all communities; however, gamma
camera PET could be in the very near future. Additional, useful
clinical information is added to patient management by these
studies and we should not allow honest disagreements over
technologic issues to impede the acquisition of vital patient
management data. AR
References
1. Cassen B: Principles of Instrumentation. In: Blahd WH (ed):
Nuclear Medicine, pp 58-60. New York, McGraw-Hill, 1965.
2. Ter-Pogossian MM, Phelps ME, Hoffman EJ, Mullani NA: A
positron-emission transaxial tomograph for nuclear imaging.
Radiology 114(1):89-98, 1975.
3. Muehllehner G, Karp JS: A positron camera using position
sensitive detectors: PENN-PET.
J Nucl Med 27(1):90-98, 1986.
4. He ZX, Scarlett MD, Mahmarian JJ, Verani MS: Enhanced
accuracy of defect detection by myocardial single-photon emission
computed tomography with attenuation correction with gadolinium-153
line sources: Evaluation with a cardiac phantom.
J Nucl Cardiol 4(3):202-210, 1997.
5. Bacharach SL, Bivat I: Attenuation correction in cardiac
positron emission tomography and single-photon emission tomography.
J Nucl Cardiol 2(3):246-255, 1995.
6. Hashimoto J, Ogawa K, Ichihara T, et al: Application of
transmission scan-based attenuation compensation to
scatter-corrected Thallium-201 myocardial single-photon emission
tomographic images. Eur J Nucl Med 25(2):120-127, 1998.
7. Valk PE, Pounds TR, Tesar RD, et al: Cost-effectiveness of
PET imaging in clinical oncology. Nucl Med Biol 23(6):737-743,
1996.
8. Henry N. Wagner, MD: Personal communication, January,
1998.
9. Grossman ZD, Lamonica MD, Klippenstein MD, et al: A
comparative study of 511 KeV SPECT and PET, for malignancy
detection using separate 370 MBq Fluorine-18-FDG on different days.
Radiology (Suppl) 205(P):221, 1997.
Dr. Henkin is Director of Nuclear Medicine and Professor of
Radiology at
Loyola University Medical Center in Maywood, IL. He is also a
member of the editorial advisory board of this journal.